Field Hydrostatic Pressure Testing Procedures - Emerson Rosemount 3410 Series Maintenance And Troubleshooting Manual

Gas ultrasonic flow meters
Hide thumbs Also See for Rosemount 3410 Series:
Table of Contents

Advertisement

Maintenance
April 2022
WARNING
CRUSHING HAZARD
Do not remove flange stabilizers.
Attempting to do so could allow the meter to roll, resulting in serious injury or equipment
damage.
A. Flange stabilizers
CAUTION
ESCAPING FLUIDS HAZARD
The purchaser of the meter is responsible for the selection of Rosemount components/
seals and materials compatible with the chemical properties of the measurement fluid.
Failure to select suitable meter components/seals may cause escaping fluids, resulting in
injury or equipment damage.
WARNING
HEARING DAMAGE
Wear proper hearing protection before approaching a metering system that is generating
a large amount of audible noise. Obey all facility safety rules.
Failure to comply could result in temporary or permanent hearing loss.
1.2

Field hydrostatic pressure testing procedures

CAUTION
LEAKAGE OR PRESSURE CONTAINING PARTS FAILURE
Use precautions to eliminate hazards to personnel in the event of leakage or failure of the
gas ultrasonic meter pressure containing parts or failure of the test equipment and to
prevent over-pressurization during the test procedure.
Failure to comply may result in injury to personnel or cause damage to the equipment.
6
Maintenance and Troubleshooting manual
Emerson.com/Rosemount
00809-0200-3104