Download Print this page

Advertisement

Quick Links

GuideSheath Kit 2 Quick Reference Guide
Single Use Guide Sheath Kit
K-401/402/403/404
Below are important steps for handling the GuideSheath Kit 2. For detailed operating instructions, be sure to
follow the instruction manual included in the device package.
1 Types of Stoppers
The GuideSheath Kit 2 comes with two types of stoppers – one for an ultrasonic probe and one for EndoTherapy
accessories.
Model
Compatible channel diameter
ET stopper
US stopper
2 Operation of ET/US Stoppers
<Preparation of US Stopper>
Insert the ultrasonic probe into the guide sheath's
1
accessory port so that the probe tip protrudes
slightly from the guide sheath tip. Attach the US
stopper to the ultrasonic probe by inserting the
probe to the end of the slit.
Note:
Make sure the ultrasonic probe
reaches the end of the slit on
the stopper.
Push the US stopper
2
connection portion
into the guide sheath's
accessory port to
secure it.
K-401
2.0 mm or more
4 pieces
(White)
1 piece
(Gray)
US stopper
3
Accessory port
K-402
K-403
3 pieces
4 pieces
(White)
(3: Gray / 1: White)
Adjust the position of the
US stopper so that the
probe tip protrudes slightly
from the guide sheath tip.
Remove the US stopper
from the guide sheath's
accessory port while
attaching the US stopper
to the ultrasonic probe.
K-404
2.8 mm or more
3 pieces
(2: Gray / 1: White)
1 piece
(Gray)

Advertisement

loading

Summary of Contents for Olympus K-401

  • Page 1 GuideSheath Kit 2 Quick Reference Guide Single Use Guide Sheath Kit K-401/402/403/404 Below are important steps for handling the GuideSheath Kit 2. For detailed operating instructions, be sure to follow the instruction manual included in the device package. 1 Types of Stoppers The GuideSheath Kit 2 comes with two types of stoppers –...
  • Page 2 As medical knowledge is constantly growing, technical modifications or changes of the product design, product specifications, accessories and service offerings may be required. Postbox 10 49 08, 20034 Hamburg, Germany Wendenstrasse 14-18, 20097 Hamburg, Germany Phone: +49 40 23773-0, Fax: +49 40 233765 www.olympus-europa.com N8602374-122021...