3.2
Machine Moving and Fixing Means
Moving Means
• Caster x 4 (omnidirectional, no brake attached)
Fixing Means
• Adjustable foot x 4
3.3

Environmental Requirements

Atmospheric Requirements
Temperature
Relative
humidity
Atmospheric
pressure
The above environmental requirements during non-operation
and during transit do not apply to IPs (Imaging Plates).
Floor (Installation Area) Vibration Requirements
10 to 55 Hz in frequency and 0.0075 mm or less in amplitude.
Floor Levelness
10 mm/m (1/100 of inclination) or less for both front and rear and both right and left
Floor Flatness
10 mm or less
Variable Magnetic Field
0.3 gauss p-p or less
009-058-00
009-058-00
10.10.2000 FM2732 (1)
10.10.2000 FM2732
Operation
Non-operation
15–30 °C
0–45 °C
10–90 %
40–80 %
(Without condensation)
700–1,030 hPa
CR-IR347 Service Manual
During transit
-10–50 °C
10–90 %
500–1,030 hPa
TR7H0001.EPS
0.17
0.17