Page 4
75023 PRODUCT INFORMATION Product Name: Concord Handheld Pulse Oximeter Product Model: 75023 Manufacturer: Concord Health Supply, Inc. Address: 9052 Terminal Ave Skokie IL 60077 Tel: 1-888-970-2999 Website:www.ConcordHealthSupply.com REVISION HISTORY This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change.
Page 5
75023 PREFACE This manual contains the instructions necessary to operate the product MANUAL PURPOSE safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety.
4. 3 Installing the Batteries ........12-13 75023 Operation ........... 13-19 5.1 lntroduction ............13 5.2 Turn On The 75023 ..........13 5.3 Set Up The 75023 ..........14-16 5.3.1 Adjust Pulse Beep Volume ......... 14 5.3.2 Select User Type ..........14 5.3.3 Set Alarm Limit ..........
75023 1 Safety Information WARNING: Explosion hazard. Do not use the 75023 in the presence of A.wARNING flammable anesthetics mixture with air, oxygen, or nitrous oxide. WARNING: Chemicals from a broken LCD display panel are toxic when inoesled. Use caution when the 75023 has a broken displav panel.
Page 9
Concord. Do not connect any device other than a Concord-approved SPO2 sensor to the sensor connector. WARNING: The 75023 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the 75023 should be observed to verify normal operation using the configuration in which ii is to be used.
Copyright. Concord Health Supply. Inc. 75023 2 Symbols Direct Current (DC) Attention: Consult accompanying documents (this manual}. � Auxiliary output connector & Power supply connector � Alarm silence button Setting button Power button Up button Down button Date of manufacture �...
Do not make any clinical judgments based solely on the 75023. The 75023 is intended only as an adjunct in patient WARNING: assessment. It must be used in conjunction with clinical signs and symptoms.
15 minutes of operation remains. The batteries should be replaced at this point. = j8PM If the 75023 is to be stored for a period of three months or longer, remove the batteries from the device before Caution: storage.
Page 13 Copyright. Concord Health Supply. Inc. 75023 WARNING: To ensure user safety, do not place the 75023 in any position that might cause it to fall on the user. WARNING: As with all medical equipment, carefully route patient cables to reduce the possibility of patient entanglement or strangulation.
75023 To access the following parameters below, press the set button until 5.3 Set up the 75023 the parameter you want to adjust is displayed. When you have completed adjusting the parameters, press and release the power button once, to save your parameters and return to the main screen.
The alarm will stay silent until the)$ key is pressed again. A_ WARNING, The 75023 will only keep the last setting. When the 75023 � is turned on, user should check whether the current selling meets conditions of user.
1) Select an appropriate SpO2 sensor according to the person category and weight. 2) Place the sensor on the appropriate measurement site. 3) Connect an approved SpO2 sensor or extension cable to the 75023 SpO2 sensor port. � �I 5.4.2 Display Description...
Page 18
. .&. Caution: Use only Accurate-approved sensors and extension cables. Physiological conditions, medical procedures, or external Note: agents that may interfere with the 75023's ability to detect and display measurements include: • Dysfunctional hemoglobin • Arterial dyes • Low perfusion •...
• Poor peripheral perfusion Note: An extension cable may be used to provide more distance between the 75023 and the sensor. Use the extension cable available from Concord Health Supply. Notifications triggered by a vital sign that appears abnormal or by technical 5.5 Alarm Notifications...
"H" means high, " M" means medium and "L" means low. The "Cause and actions" column gives recommendations to instruct you to troubleshoot the problems. If the problem persists, contact your service personnel. Copyright. Concord Health Supply. Inc. Page 19...
The screen will show type are correct, and make "OFF" sure that the sensor is undamaged. Reconnect the sensor if the sensor is disconnected or use a new sensor if the sensor is damaged, for example. Page 20 Copyright. Concord Health Supply. Inc.
2 ) The data from the hypoxia b reathe-d own trials for each of the 1 2 volu n teers i s summari zed i n F i g u res 1 &2 : Copyright. Concord Health Supply. Inc. Page21...
Page 23
75023 -• .§ <( "C � • • -• Page 22 Copyright. Concord Health Supply. Inc.
Page 24
75023 "iii '" :::; ll:'. � � • • Page 23 Copyright. Concord Health Supply. Inc.
70-100% and the pulse rate accuracy is less than ±3 in range 25-250 bpm, which meets the requirements of ISO80601-2-61. 7 M aintenance WARNING: Do not spray, pour, or spill any liquid on the 75023, its 7.1 Cleaning accessories, connectors, switches, or openings in the chassis as this ma dama e the pulse oximeter.
(more frequently if perfusion is poor). • Do not immerse or wet the sensor. • Do not lift the 75023 by the sensor or the extension cable because the cable could disconnect from the 75023,causing the 75023 to drop on the...
Copyright. Concord Health Supply. Inc. 75023 Testing Intended Information Of Sensor Model Picture Name position crowd U403-01 Reusable Adult Finger Adult Finger Clip SpO2 (�40kg) Sensor 1) Hold the sensor with its opening towards the user's index or ring Sensor Usager finger (Fig.
Page 28
Copyright. Concord Health Supply. Inc. 75023 C o m m u n i cation port D a t a storage S h e l f l ife 5 Yea rs Al kal lne batta rlea Q u a ntity Specification 1 .
CISPR 1 1 cause any interference in nearby electronic equipment. Radio frequency Class A The 75023 is suitable for use in all (RF) emissions establishments other than domestic and CISPR 1 1 those directly connected to the public Not applicable low-voltage power supply network that...
Page 30
UT for0.5 applicable be that of a typical commercial short interruptions cycles or hospital environment. If the user of the 75023 requires and voltage 40 % UT (60 % dip continued operation during power supply in UT) for 5 cycles power mains interru �...
Page 31
Page30 Copyright. Concord Health Supply. Inc. 75023 Note 1: From 80 MHz to 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.