Technical Information
When the Request is "E, the Answer will contain the complete set of the following
parameters:
Table 15
Bytes
2÷5
6÷9
10÷11
12
13
14÷17
18÷19
20
21
22
* The values for current and voltage are given as numbers, scaled from 0 to 255
where 0 corresponds to 0 V and 255 to the full scale voltage (130 V) or current
(2.5 A).
Agilent TwisTorr 305 FS Remote Controller User Manual
Meaning
1
MSB
LSB
- - X X X X
0
= STOP
1
= WAITING INTERLOCK
2
= STARTING
3
= NORMAL OPERATION
4, 5 = HIGH LOAD
6
= FAILURE
7 = APPROACHING LOW SPEED
Cycle Time
Pump Life
Pump Temperature
Current*
Voltage*
Frequency
Cycle #
R1 Status
R2 Status
CRC