ResMed AirSense 11 User Manual page 26

Hide thumbs Also See for AirSense 11:
Table of Contents

Advertisement

Electromagnetic compatibility
The AirSense 11 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC 60601-1-2:2020,
for residential, commercial and light industry environments.
Portable and mobile RF communications equipment should be used no closer to any part of the machine, including cables, than the
recommended 3.94" (10 cm) separation distance.
The AirSense 11 has been designed to meet EMC standards. However, should you suspect that the device performance (eg.
pressure or flow) is affected by other equipment, move the device away from the possible cause of interference.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found in
ResMed.com/downloads/devices.
IEC 60601-1 (Edition 3.1) classification
Class II (double insulation), Type BF, Ingress protection IP22.
Supplemental oxygen maximum flow
15 L/min
Aircraft use
ResMed confirms that the machine meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M; RTCA-DO-160, section 20, category T) for all phases of air travel.
Design life
Device, power supply unit:
Standard humidifier tub:
Cleanable humidifier tub
Air tubing:
General
The patient is an intended operator.
Pneumatic flow path
Displayed values
Value
Pressure at mask:
Displayed mask pressure
Flow derived values:
Leak
1
1
Results may be inaccurate in the presence of leaks or supplemental oxygen
Pressure accuracy
Maximum static pressure variation at 10 cm H
Device with HumidAir 11 tub and air tubing:
Device with Side cover and air tubing
Note: Refer to the relevant measurement uncertainty from the Measurement system uncertainties table.
Range
1
4-20 cm H
O (4-20 hPa)
2
0-120 L/min
5 years
6 months
2.5 years
6 months
1.
Flow sensor
2.
Blower
3.
Pressure sensor
4.
Mask
5.
Air tubing
6.
Humidifier
7.
Device
8.
Inlet filter
Accuracy
±0.5 cm H
O (0.5 hPa) ±4% of
2
measured reading
± 12 L/min or 20% of reading
whichever is greater, 0 to 60
L/min
O (10 hPa) according to ISO 80601-2-70:2015
2
±0.5 cm H
O (±0.5 hPa)
2
±0.5 cm H
O (±0.5 hPa)
2
Display resolution
0.1 cm H
O (0.1 hPa)
2
1 L/min
English
23