Using the SCC-A10 when Scaling Voltage Measurements

Specifications

In order for the SCC-A10 to perform according to the specifications in this guide,
Note
the ±15 V power supply you use must be accurate to within 5%.

Analog Input

Temperature range is 23 °C ± 5 °C.
1
2
Calculated with respect to the input range
© National Instruments Corporation
If you configured the SCC-A10 using Measurement & Automation
Explorer (MAX) and you are using NI-DAQ, the voltage reading you get
from the E Series DAQ device accounts for the voltage scaling effect of the
SCC-A10. Otherwise, because the voltage measurement from the E Series
DAQ device is one-tenth of the voltage applied at the SCC-A10 input, you
must multiply the voltage reading from the E Series DAQ device by 10 to
get the correct SCC-A10 input voltage.
These ratings are typical at 25 °C unless otherwise stated.
Number of input channels ...................... 2 DIFF
Input range ............................................. ±100 VDC
Input impedance..................................... 1 MΩ (powered on or off)
Gain error ............................................... ±0.14% of reading max
Gain-error temperature coefficient......... ±0.006% of reading/°C max
Offset error............................................. ±6.5 mV max (referred to
Offset-error temperature coefficient ...... ±80 µV/°C max (RTI)
Nonlinearity ........................................... ±0.004% of full scale
Common-mode rejection ratio ............... 60 dB min (DC to 60 Hz)
input [RTI]
5
SCC-A10 Voltage Attenuator Module User Guide
1