Re-Programming Ph Measurement - Hach sensION+ PH 31 Manual

Hide thumbs Also See for sensION+ PH 31:
Table of Contents

Advertisement

4. Operation

4.8. Re-programming pH measurement

Default programming (standard)
Resolution
• 0.01 pH
Measuring mode
• By stability .
- Standard criterion (variation < 0 .01 pH in 6 s) .
• In continuous .
- The result is stored when the measurement
has finished .
- The result is printed when the measurement
has finished .
View mV
• No
Limits
• No
pH isopotencial
• pHiso 7.00
"Re-programming" flow chart
10-04-2010
17:15
pH MEASURE
New calibration
in 1 day
...by stability
Measurement condit.
Resolution: 0.01
...
Meas.: By stability
Start Measurement
Display mV: No
Limits: No
Isopotential: pH 7
Start measurement
Measurement condit.
Resolution: 0.01
Meas.: By stability
Display mV: No
Limits: No
Isopotential: pH 7
Measurement condit.
Resolution: 0.01
Meas.: By stability
Display mV: No
Limits: No
Isopotential: pH 7
Measurement condit.
Resolution: 0.01
Meas.: By stability
Display mV: No
Limits: No
Isopotential: pH 7
Measurement condit.
Resolution: 0.01
Meas.: By stability
Start of
Display mV: No
Limits: No
measurement with
Isopotential: pH 7
new programming.
etc.
Other options
• 0.1 pH
• By stability.
- Rapid criterion (variation < 0 .02 pH in 6 s) .
- Standard criterion (variation < 0 .01 pH in 6 s) .
- Strict criterion (variation < 0 .002 pH in 6 s) .
• In continuous . Selection of the time interval for:
- Data storage .
- Data printing .
• By time . Selection of the time interval for:
- Data storage .
- Data printing .
• Simultaneous display of pH and mV concerned.
• Higher and lower limits. Acoustic warning when the measure-
ment is out of limits, in the printed ticket appears "A" beside
the measured value .
pHiso 7 .00 / Manual introduction / Value calculation .
Important note:
Once the measurement has been re-programmed, it will be performed in
accordance with the new parameters .
It is highly recommended to apply the same stability criteria to measure and
calibrate .
Measurement condit.
Resolution pH:
0.1
0.01
0.001
Measurement condit.
Measurement condit.
Meas.:
By Stability
By stability
Criteria
In Continuous
Rapid
By time
Standard
Strict
Measurement condit.
Measurement condit.
Meas.:
In Continuous
By stability
In Continuous
Acquis. Interval
By time
01:00
Measurement condit.
Measurement condit.
Meas.:
By time
By stability
In Continuous
Measurement time
By time
00:01:00
Measurement condit.
Display mV:
Yes
No
Measurement condit.
Measurement condit.
Limits:
Limits
Yes
Upper: pH 8.20
No
Lower: pH 6.05
Exit
Measurement condit.
Isopotencial:
pH 7
See Electrode isopotential on page 18.
Data introduction
Calculate
• 0.01 pH
• 0.001 pH
Measurement condit.
By Stability
Criteria
Rapid
Standard
Strict
Measurement condit.
In Continuous
Acquis. Interval
Data storage in
01:30
the Data Logger. It
coincides with the
printing and PC
Measurement condit.
sending. Selecting
By time
0 store/print at
Acquis. Interval
the end of the
01:30
measurement.
Measurement condit.
Lower= pH 6.05
Upper limit?
pH
8.20
11

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents