2) Buzzer Alarm
a.The buzzer beeps once (about 0.25s) at any valid press of buttons or turning of the function dial.
b.The buzzer beeps continuously when the input voltage or current is over range.
3) Low Battery Detection
When the battery voltage is <4.6V, "
" is displayed. After the symbol lasts for 20s, the meter
displays "Lo.bt" and automatically shuts down after 60s.
IX. Specifications
1. General Specifications
1) Max display: 19999
2) Polarity display: Auto
3) Overload display: "OL" or "-OL"
4) Low battery indication: "
" is displayed.
5) Battery: 4×1.5V AAA
6) Auto power off: If there is no operation of the function dial or any button for 15 minutes, the meter
will automatically power off. This function can be disabled.
7) Dimensions: 189.4mm×89mm×53.8mm
8) Weight: About 370g (including batteries)
9) Operating altitude: 2000m
10) Operating temperature and humidity: 0°C~30°C (≤80%RH), 30°C~40°C (≤75%RH),
40°C~50°C (≤45%RH)
11) Storage temperature and humidity: -20°C~60°C (≤80%RH)
12) Electromagnetic compatibility:
RF=1V/m, overall accuracy = specified accuracy + 5% of range
RF>1V/m, no specified calculation
2. Electrical Specifications
Accuracy: ± (a% of reading + b digits), 1 year calibration period
Ambient temperature: 23°C ± 5°C; ≤80%RH
Temperature coefficient: To ensure measurement accuracy, the operating temperature should be
within 18°C~28°C and the fluctuation range should be within ±1°C. When the temperature is <18°C
or >28°C, add temperature coefficient error: 0.1 x (specified accuracy)/°C.
1) AC Voltage
Range
Resolution
200.00mV
0.01mV
2.0000V
0.0001V
20.000V
0.001V
200.00V
0.01V
1000V
1V
Overload protection: 1000V
Input impedance: About 10MΩ
Frequency response: 45Hz~400Hz, true RMS display
Accuracy guarantee: 5%~100% of range
AC crest factor: The full-scale value of the AC crest factor can reach 3.0. If the crest factor of a
non-sinusoidal wave is ≤3.0, no additional error will be added. The accuracy of other crest factors
beyond the range are not assessed.
2) LPF AC Voltage
Range
Resolution
200.00V
0.01V
1000.0V
0.1V
Overload protection: 1000V
LPF reading = measured value × 2 ± (2%+30)
3) DC Voltage
Range
Resolution
200.00mV
0.01mV
2.0000V
0.0001V
20.000V
0.001V
200.00V
0.01V
1000V
1V
Overload protection: 1000V
Input impedance: About 10MΩ
For mV range, short circuit allows least significant digit ≤5.
Accuracy guarantee: 1%~100% of range
4) AC Current
Range
Resolution
2000.0µA
0.1 A
µ
200.00mA
0.01mA
20.000A
0.001A
Overload protection: 250Vrms
Frequency response: 45Hz~400Hz, true RMS display
AC crest factor: The full-scale value of the AC crest factor can reach 3.0. If the crest factor of a
non-sinusoidal wave is ≤3.0, no additional error will be added. The accuracy of other crest factors
beyond the range are not assessed.
Accuracy guarantee:5%~100% of range.
5) DC Current
Range
Resolution
200.00 A
µ
0.01 A
µ
2000.0 A
µ
0.1 A
µ
200.00mA
0.01mA
20.000A
0.001A
Overload protection: 250Vrms
Accuracy guarantee:1%~100% of range.
The actual reading shall increase 3 digits if lower than 5% of range.
6) Resistance
Range
200.00Ω
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
20.000MΩ
200.0MΩ
Overload protection: 600V
7) Conductance
Range
200.00nS
Overload protection: 600V
8) Continuity
Range
200.00Ω
Overload protection: 600V
9) Diode
Range
Accuracy
20.000V
±(1.0%+25)
Overload protection: 600V
±(0.8%+25)
10) Capacitance
±(
)
Range
1.2%+25
20.000nF
200.00nF
2000.0nF
2.0000 F
20.000 F
200.00 F
2.0000mF
20.000mF
Accuracy
200.00mF
±(2%+30)
Overload protection: 600V
Measurement result = displayed value – capacitance of open-circuit test leads
For capacitance ≤100nF, it is recommended to use the relative value mode.
Open circuit allows least significant digit ≤20.
11) Temperature
Range
Accuracy
-40℃~40℃
40℃~400℃
400℃~1000℃
±(0.05%+5)
-40℉~104℉
104℉~752℉
752℉~1832℉
±(0.1%+5)
Overload protection:600V
12 Frequency
Range
200.00Hz 60.00MHz
Overload protection:600V
Accuracy
X. Maintenance
Warning: Before opening the rear cover or battery cover,
±(1.5%+15)
switch off the power supply and remove the test leads.
1. General Maintenance
±(2.5%+35)
1) Clean the meter casing with a soft cloth and mild detergent.
Do not use abrasives or solvents!
2) If there is any malfunction, stop using the meter and send it
for maintenance.
3) The maintenance and service must be implemented by
qualified professionals or designated departments.
2. Battery/Fuse Replacement
1) Turn the function dial to the "OFF" position, and remove the test leads from the input terminals.
2) Unscrew and remove the battery cover.
Accuracy
3) Replace with 4×1.5V AAA batteries, observing correct polarity.
Or replace the blown fuse (specifications: F1 Fuse 630mA 250V Φ5x20mm ceramic tube;
±(0.5%+4)
F2 Fuse 20A 250V Φ5x20mm ceramic tube)
4) Secure the battery cover and tighten the screw.
±(0.8%+6)
±(2.0%+15)
Resolution
0.01Ω
0.0001kΩ
0.001kΩ
0.01kΩ
0.0001MΩ
0.001MΩ
0.1MΩ
Resolution
0.01nS
Resolution
0.01Ω
Resolution
Open circuit voltage: About 2.8V
Measurable PN junction: Forward
0.001V
voltage drop ≤2V
For silicon PN junction, the normal
value is generally about 0.5V~0.8V.
Resolution
0.001nF
0.01nF
0.1nF
0.0001 F
µ
µ
0.001 F
µ
µ
0.01 F
µ
µ
0.1 F
µ
0.001mF
0.01mF
Resolution
0.1℃
0.1℉
Resolution
0.01Hz 0.01MHz
Accuracy
±(0.4%+10)
±(0.4%+5)
±(1.2%+25)
±(5.0%+10)
Accuracy
±(1.2%+50)
Accuracy
<10Ω: Consecutive beeps
≥50Ω: No beep
Open circuit voltage: About 2.8V
Accuracy
Accuracy
±(4.0%+50)
±(10%)
Accuracy
±4℃
±(1.0%+5 )
±(2.0%+
5℃
)
±5℉
±(1.5%+
5℉
)
±(2.5%+
5℉
)
Accuracy
±(
)
0.1%+3
Screw
Battery
Fuse