Fluke 199XRAY User Manual

Fluke 199XRAY User Manual

Medical scopemeter

Advertisement

Quick Links

®
Fluke199XRAY
Medical ScopeMeter
Users Manual
4822 872 30791
August 2006
© 2006 Fluke Corporation, All rights reserved.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 199XRAY

  • Page 1 ® Fluke199XRAY Medical ScopeMeter Users Manual 4822 872 30791 August 2006 © 2006 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies.
  • Page 3: Table Of Contents

    Safety Information ......................1-1 Limited Warranty, Limitation of Liability ................1-1 Shipment Contents......................1-1 About the X-RAY kVp Function ..................1-2 190B/C vs. 199XRAY User Interface................1-3 USER Button Bar....................... 1-3 190B/C RECORDER Key ..................1-4 199XRAY MEDICAL Key................... 1-4 kVp Measurement Procedure..................
  • Page 4 Fluke199XRAY Users Manual Fluoro Mode (FLUORO).................... 2-7 kV Measurements ..................... 2-7 Exposure Time Measurements (RADIO only) ............2-8 AUTO-kVp Mode, MANUAL Mode ................2-9 Selecting AUTO-kVp mode – Manual Mode............2-9 Auto kVp Mode Waveform Requirements ............2-9 kVp Scope Mode Screen, Readings, F1-F4 Keys ............. 2-10 kVp Mode Readings .....................
  • Page 5: About This Manual

    Fluke 190B/C and the Fluke 199XRAY. Shipment Contents For other information on the Fluke 199XRAY please In addition to the shipment contents shown in the Fluke consult the Fluke 190B/C Series Getting Started manual, 192B-196B/C-199B/C Users Manual, the 199XRAY...
  • Page 6: About The X-Ray Kvp Function

    About the X-RAY kVp Function The X-RAY kVp function provides oscilloscope setups for measurements using the kVp Divider and the following The 199XRAY Medical Scopemeter, when used with Fluke filter packs on Input A : Biomedical kVp Divider, automatically measure kVp •...
  • Page 7: 190B/C Vs. 199Xray User Interface

    Pressing the 190B/C USER key will open equal, except for the following: the button bar: • The 199XRAY user interface text is available in English only, language selection is not possible. See ‘USER Button Bar ’ below. Pressing the 199XRAY USER key will open the button bar without the •...
  • Page 8: 190B/C Recorder Key

    Fluke 199XRAY Users Manual 190B/C RECORDER Key 199XRAY MEDICAL Key Pressing the RECORDER key will open Pressing the MEDICAL key will open the screen shown below. the screen below. Press again or press to exit. The Medical measurements kVp function is described in Chapter 2 of this manual.
  • Page 9: Kvp Measurement Procedure

    Select ■ kVp with the arrow keys, Connect the kVp divider to Input A of the then accept with the F4 (ENTER) 199XRAY using a shielded 1:1 cable, for key. One of the Filter Pack example the cable of the MA190 Accessory set.
  • Page 10: Wide Range - Low Range Filter Pack

    Fluke 199XRAY Users Manual Wide Range – Low Range Filter Pack CT & Mobile Pack Plus Filter Pack Similar screens are shown for Wide Range and Low Similar screens are shown for Wide Range and Low Range. Range. Proceed as follows to select the filter pack: Proceed as follows to select the filter pack: Select ■...
  • Page 11: Cadmium K-Edge Mammo Filter Pack Calibration Overview

    Using the X-RAY kVp Function Selecting a Filter Pack 25 kVp Measurement Procedure: Cadmium K-edge Mammo Filter Pack Calibration Overview Set the x-ray generator for 25 kVp, 100 mA, 80 ms. If not During Cadmium K-edge Mammo Filter Pack calibration, available, use a mAs setting that provides an exposure of screen instructions will guide you through the calibration 80 ms (+/- 10 ms).
  • Page 12 Fluke 199XRAY Users Manual If the exposure is within the range 27.5 to 29.5 kVp, the Linear Correction Exposure: reading is accurate. If ---.kVp is displayed, set the Make an exposure using the same generator settings as generator for 28 kVp and repeat step 4. If 0.0 kVp is those used to make the last k-edge exposure.
  • Page 13: K-Edge Mammo Filter Pack

    Using the X-RAY kVp Function Selecting a Filter Pack k-Edge Mammo Filter Pack Accept ■ Recalibrate… to start the calibration. The k-Edge Mammo filter pack is used to determine the correction factor (recalibrate) for the Linear Mammo filter You will see the calibration startup pack.
  • Page 14: Linear Mammo Filter Pack

    Fluke 199XRAY Users Manual Linear Mammo Filter Pack Select and accept: ■ Recalibrate… to determine the correction factor for the linear Mammo filter pack. Continue as for the k-Edge Mammo filter pack step 3. ■ Use cal. values… to use the present calibration values.
  • Page 15: Kvp Scope Mode

    Scope Mode Fluoro Mode (FLUORO) kVp Scope Mode When used in the Fluoro mode, the 199XRAY waits for a When entering the kVp Scope mode predefined scope trigger then samples the kV waveform at one sample per settings for the selected filter pack are provided. Some of second.
  • Page 16: Exposure Time Measurements (Radio Only)

    Fluke 199XRAY Users Manual AUTO-kVp Mode, MANUAL Mode Note: When measuring kV, the bottom of the trace In the AUTO-kVp mode, vertical cursor lines are (indicated by the A input zero icon “A-“ at the left automatically positioned to indicate the area used for the of the screen) is not zero.
  • Page 17: Auto Kvp Mode Waveform Requirements

    Using the X-RAY kVp Function kVp Scope Mode For automatic measurement of kV and exposure time in Auto kVp Mode Waveform Requirements the radiographic measurement mode, the entire waveform For automatic measurement of kV in the Fluoroscopic must be onscreen and must meet the following criteria: measurement mode, the waveform must be onscreen and must meet the following criteria: •...
  • Page 18: Kvp Scope Mode Screen, Readings, F1-F4 Keys

    Fluke 199XRAY Users Manual In the MANUAL mode other readings may be shown kVp Scope Mode Screen, Readings, F1-F4 Keys depending on the selected cursors. See ‘F1…F4 softkey An example of the kVp scope mode screen is shown in the functions for cursor menu ON’...
  • Page 19 Using the X-RAY kVp Function kVp Scope Mode F1…F4 Softkey Functions MANUAL mode: Two vertical cursors. Use to move the cursors. Use F2 to select F1…F4 softkey functions for cursor menu OFF: cursor to be moved. The key label shows the current filter pack name.
  • Page 20: Kvp Scope Mode Presets

    Fluke 199XRAY Users Manual Sensitivity, Probe Factor, and Bandwidth filter are kVp Scope Mode Presets fixed. When entering the kVp mode some Scope functions are preset. Some of these function setting are fixed, some can Note: be changed. See Table 2-1 and Table 2-2 for details.
  • Page 21 Using the X-RAY kVp Function kVp Scope Mode Table 2-2. kV Range and Trigger Level vs. Filter Pack and Vertical Sensitivity Filter Pack Vertical Sensitivity kV Range Trigger Level Wide Range 20 kV/div * 80 -150 kV 80 kV 10 kV/div 60 –...
  • Page 22: Specifications

    Fluke 199XRAY Users Manual Specifications The 199XRAY kVp Measurement Accuracy Specifications are shown in table 2-3. Table 2-3. 199XRAY kVp Measurement Accurcy Specifications Filter Pack kVp-Avg Accuracy kV Peak Accuracy* Exposure Time Accuracy Large Signal Small Signal Wide-Range 1.5% 0.5% K-Edge Mammo 0.5%...