Step 4: BC 85
data transfer immediately
following measurements.
®
• If Bluetooth
data trans-
fer has been activated,
data is transferred after
having confirmed the user
memory and pressing the
START/STOP button
The "beurer HealthManager" app must be active to allow
data transfers.
If your smartphone has a protective cover, remove this to en-
sure that there is no interference during the transfer. Begin the
data transfer in the "beurer HealthManager" app.

9. Error messages/troubleshooting

In the event of errors, the error message E_ appears on the
display.
Error messages may appear if:
• It was not possible to record the pulse: E 1;
• You move or speak during the measurement:: E2 ;
• The cuff is fastened too tightly or loosely: E3;
• Errors occur during the measurement: E4;
• The pump pressure is higher than 300 mmHg: E5;
• The batteries are almost empty
• The data could not be sent via Bluetooth
Step 4: BC 85
Data transfer at a later point:
• Go to memory mode
(chapter 7). Select the
desired user memory. The
®
Bluetooth
transfer starts
automatically.
.
E6.
®
:
.
In such cases, repeat the measurement. Ensure that you do not
move or speak during the measurement. If necessary, reinsert
or replace the batteries.

10. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries.

11. Technical specifications

Model
Type
Measurement method Oscillometric, non-invasive blood pres-
Measurement range
Display accuracy
Measurement inac-
curacy
30
BC 85
BC 80
sure measurement on the wrist
Cuff pressure 0-300 mmHg,
Systolic 50-250 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-180 beats/minute
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg