General
1.14

Technical Safety Checks (TSC)

Abbreviation: TSC = Technical Safety Checks
NOTE
Technical Safety Checks,
Test certificate - check
1 Identification data
2 General checks
2.1 Visual inspection: Is there any damage
affecting safety?
2.2 Are all cables and cable guides firmly
attached and in place, without any visible
damage?
Siemens AG
Medical Solutions
Within the purview of DIN VDE 0751-1, the operator of medical
engineering products has to perform technical safety checks at
regular intervals.
The checks listed in these maintenance instructions contain all
technical safety checks according to DIN VDE 0751-1 – except
for the checks regarding the completeness, presence and legi-
bility of the necessary operator documents.
A separate protocol with print number SP00.000.834.01... is
available for each technical safety check.
Within the purview of DIN VDE 0751-1, we recommend docu-
menting the maintenance results both in the maintenance pro-
tocol and in the TSC protocol. The protocols should be filled
out completely and handed over to the client after maintenance
is completed.
In the table below, each technical safety check listed in docu-
ment SP00.000.834.01... is assigned to the corresponding
maintenance work.
SPR8-215.831.04.03.02
07.05
Maintenance instructions - check/
Comments
Please fill in the fields.
n.a.
2.1.1 Damage, checks:
SIM Covers
SIM Cassette compartment
SIM Control and display panel
SIM Exposure release cable
SIM Single tank generator
SIM Collimator
2.2.10 Mains cable, checks:
SIM Cable winder
SIM Damage
SIM Protective conductor
2.2.12 Arm cable harness, check:
SIM Damage
Page 19 of 64