HP 85091A Reference Manual page 26

Electronic calibration modules
Table of Contents

Advertisement

Table 2-6
Type-N 75
Device
Type-N 75 Ω ECal Module
Type-N 75 Ω Adapter
1. Measured from a 5.258 mm (0.207 inches) nominal offset.
2. Approximately +2 sigma to –2 sigma of gauge uncertainty based on studies performed at the factory using analog gauges
according to recommended procedures.
Chapter 2
Connector Pin Depth Limits
1
Typical Pin Depth
in
–4
micrometers (10
inches)
–25.4 to –50.8 (–10.0 to –20.0) +3.8 to –3.8 (+1.5 to –1.5)
0 to –76.2 (0 to –30.0)
Measurement Uncertainty
–4
in micrometers (10
inches)
+3.8 to –3.8 (+1.5 to –1.5)
Specifications and Characteristics
Connector Pin Depth Information
2
Observed Pin Depth Limits
–4
in micrometers (10
inches)
–21.6 to –54.6 (–8.5 to –21.5)
+3.8 to –80.0 (+1.5 to –31.5)
2-9

Advertisement

Table of Contents
loading

Table of Contents