  • Page 1 VistOR Installation & Service Manual...
  • Page 2 Luminaries and lumenaires for diagnosis In accordance with IEC-60601-2-41 Intended use The VistOR Series are AC powered devices that provide a field of illumination for general examination and surgery. Page 2 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410...
  • Page 3: Table Of Contents

    Section 1: Terminology Definition of Terms List of Symbols Section 2: Specifications VistOR PR Specifications VistOR PRO Specifications VistOR SP and VistOR SPF Specifications VistOR EX & EXL Specifications Chuttle Track Specifications Warranty Section 3: Installation/Assembly Ceiling Mount Pre-Installation Guidelines...
  • Page 4 Maintenance Schedule Section 7: Troubleshooting General Troubleshooting Section 8: Exploded Views and Parts List VistOR PR, VistOR PRO and VistOR EX & EXL Electric Schematic VistOR SP Electric Schematic VistOR SPF Electric Schematic VistOR PR Exploded View, Light Head Assembly VistOR PR Light Head Assembly Part’s List...
  • Page 5 Figure 32: VistOR SP Spotlight Operation Figure 33: VistOR SP Spotlight Controls Figure 34: VistOR SPF Spotlight Model Operation Figure 35: VistOR EX & EXL Diagnostic/Specialty Light Operation Figure 36: VistOR PR/VistOR PRO Lamp Replacement Figure 37: VistOR EX & EXL Lamp Replacement Page 5 of 85 5368 KUHL ROAD ●...
  • Page 6 Figure 44: Electric Schematic for VistOR PR & PRO and VistOR EX & EXL, 120VAC, 230 VAC, &240 VAC Figure 45: Electric Schematic for VistOR PR & PRO & VistOR EX & EXL, 100VAC Figure 46: Electric Schematic for VistOR SP, 120VAC & 230 VAC...
  • Page 7: Section 1: Terminology

    Section 1 Terminology Definition of Terms I.E.C. International Electrotechnical Commission U.L. Underwriter’s Laboratories Medical Electrical Equipment Electrical equipment intended to diagnose, treat the patient under medical supervision. Electrical equipment that transfers energy to the patient . Central Illuminance Illuminance of light head measured at 1 meter from the light emitting area with no obstructions.
  • Page 8 Section 1 Definition of Terms (continued) Light Head/Articulating Arm Assembly: That part of the device that includes the light source, heat removal system, and light head vertical positioning arm. Extension Arm Horizontal section of the positioning arm with pivots on both ends that is used to increase the area covered by the light head and articulating arm.
  • Page 9: List Of Symbols

    Section 1 List of Symbols U.L. Listing Marking Read accompanying documents C.E. Marking Protective Earth Ground Caution Electric Shock Hazard Fuse Marking Neutral Conductor Page 9 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 10: Section 2: Specifications

    Section 2 Specifications VistOR PR Specifications Mechanical Parameter Value Weight Head Approximately 9.10 lbs. (4.14 kg). Approximately 12.5 lbs. (5.68 kg). Arm (Floor) Approximately 11.4 lbs. (5.18 kg). Wall Bracket Assy. Approximately 3.00 lbs. (1.36 kg). Floor Stand Assy. Approximately 43.0 lbs. (19.50 kg).
  • Page 11: Vistor Pro Specifications

    Section 2 VistOR PRO Specifications Mechanical Parameter Value Weight Head Approximately 10.10 lbs. (4.58 kg). Approximately 12.5 lbs. (5.68 kg). Arm (Floor) Approximately 11.4 lbs. (5.18 kg). Wall Bracket Assy. Approximately 3.00 lbs. (1.36 kg). Floor Stand Assy. Approximately 43.0 lbs. (19.50 kg).
  • Page 12: Vistor Sp And Vistor Spf Specifications

    Section 2 VistOR SP and VistOR SPF Specifications Mechanical Parameter Value Weight Head (Std.) Approximately 7.20 lbs. (3.27 kg). Head (Std. F/O) Approximately 8.00 lbs. (3.64 kg). Approximately 13.3 lbs (6.05 kg). Arm (Fiber Optic) Approximately 15.2 lbs. (6.91 kg).
  • Page 13: Vistor Ex & Exl Specifications

    Section 2 VistOR EX & EXL Specifications Mechanical Parameter Value Weight Head Approximately 3.50 lbs. (1.59 kg). Approximately 12.5 lbs. (5.68 kg). Arm (Floor) Approximately 11.4 lbs. (5.18 kg). Wall Bracket Assy. Approximately 3.00 lbs. (1.36 kg). Floor Stand Assy.
  • Page 14: Chuttle Track Specifications

    Section 2 Chuttle Track Specifications Mechanical Parameter Value Weight Single Track Assembly Approximately 44.00 lbs. (19.95 kg). Dual Trolley Model Approximately 48.00 lbs. (21.77 kg). Dimensions Track (standard) 72” (182.88 cm) Long X 10 ½” Wide (26.67 cm) X 5” (12.7 cm) High Electrical Power cable exiting the track can be connected to 115/120 VAC 60 Hz.
  • Page 15 Section 2 Page 15 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 16: Section 3: Installation/Assembly

    Ceiling Mount Pre-Installation Guidelines SPECIAL NOTE: Installation and repair of this equipment should be performed by qualified persons only. Nuvo, Inc. does not warranty any damage occurring as a result of improper installation. It is recommended that this installation manual be completely reviewed prior to installation.
  • Page 17: Ceiling Rod Calculation Single Ceiling Mount

    Rod Calculation instructions. Note: The VistOR EXL Diagnostic / Specialty Light is shown for reference. The dimensions indicated also apply to the VistOR PR Diagnostic, VistOR PRO minor surgery light and the VistOR SP Spot light. Dimensions are approximate. *Ceiling Rod Table applies to single and dual ceiling mounts.
  • Page 18: Ceiling Rod Calculation Dual Ceiling Mount

    Mount Ceiling Rod Calculation Note: The VistOR PR / VistOR SP Combination Light is shown for reference. The dimensions indicated also apply to all other combination of fixtures and dual mounting of the same light. Dimensions are approximate *Length measured when Arm is in horizontal position.
  • Page 19: Extended Ceiling Mount

    Section 3 Extended Ceiling Mount Use the following table to select the correct length ceiling rod for your application. Extended Ceiling Ceiling Rod Mounting Height Extension P/N Mount Kit P/N Extension Length 1000976-49 20.0” (508,0 mm) 10’ 6”-11’ 8” (3,20-3,56 m) 1001263 1000976-63 34.0”...
  • Page 20: Extended Ceiling Mount Installation

    Section 3 Extended Ceiling Mount Installation For 10’6”–14’ (3.2m–4.27m) Ceilings • Insert the ceiling rod up through the ceiling casting until the hole in the ceiling rod becomes visible. Insert the dowel pin into the hole in the ceiling rod and lower the rod in the ceiling casting, making sure the pin is seated securely in the indentation on the ceiling casting.
  • Page 21: Single Ceiling Mount Installation

    Installation and repair of this equipment should be performed by qualified persons only. Nuvo does not warranty any damage occurring as a result of improper installation. The shipping cartons contain a light head assembly, an arm assembly, a ceiling casting, casting cover, collar, bolt cover, ceiling rod, hardware kit, wire harness, and an Installation and Service Manual.
  • Page 22 Section 3 Single Ceiling Mount Installation • Insert the ceiling rod up through the ceiling casting until the hole in the ceiling rod becomes visible (see figure 6). Insert the dowel pin into the hole in the ceiling rod and lower the rod in the ceiling casting, making sure the pin is seated securely in the indentation on the ceiling casting.
  • Page 23: Figure 5: Ceiling Casting Mounting Diagram

    Section 3 Single Ceiling Mount Installation • Raise the extension arm to the ceiling rod and plug the Molex connectors together. Push the wire into the ceiling rod while inserting the transformer housing into the rod. Secure the housing to the ceiling rod using the four supplied bolts as shown. Two ¼”-20 x ¼” screws must be installed on one side of the ceiling rod.
  • Page 24: Figure 6: Single Ceiling Mount Installation

    Section 3 Single Ceiling Mount Installation Wire Harness (1) P/N 1000159 Dowel Pin (1) Ceiling Casting (1) P/N 0001553 P/N 1001075 Mounting Hardware (NOT PROVIDED) ” DIA Bolt (4) Set Screw (2) P/N 0001128 Casting Cover P/N 1001226 Ceiling Rod (1) P/N 16”- 1001186 P/N 25”- 1001187 P/N 35”- 1001188...
  • Page 25: Dual Ceiling Mount Installation

    Important: thoroughly check each box for parts that may be located in areas that can be overlooked. Note: The VistOR PR / VistOR SP Combination Light is shown for reference. Page 25 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410...
  • Page 26 Section 3 Dual Ceiling Mount Installation • Insert the ceiling rod up through the ceiling casting until the hole in the ceiling rod becomes visible (see figure 10). Insert the dowel pin into the hole in the ceiling rod and lower the rod in the ceiling casting, making sure the pin is seated securely in the indentation on the ceiling casting.
  • Page 27: Figure 8: Wire Harness Positioning

    Section 3 Dual Ceiling Mount Installation • Bring one section of the dual ceiling mount casting to the ceiling rod and fasten it to the rod with the 2” long bolts (P/N 0001311) and cap nuts (P/N 0001312). Tighten the cap nuts just enough to hold the section in place;...
  • Page 28: Figure 10: Dual Ceiling Mount Installation

    Section 3 Dual Ceiling Mount Installation Ceiling Rod (1) Wire Harness (1) P/N 16” 10001186 P/N 1000160 P/N 25” 10001187 P/N 35” 10001188 Dowel Pin (1) Ceiling Casting (1) P/N 0001553 P/N 1001075 Mounting Set Screw (2) Hardware P/N 0001128 (Not Provided) Casting Cover (1) Collar (1)
  • Page 29 Section 3 Dual Ceiling Mount Installation • Position the two extension arms as shown in figure 10 to allow for easier mounting of the remaining Dual mount section. Firmly press the fastened dual ceiling mount section against the ceiling rod. While firmly holding the Dual ceiling mount section in place, remove only the cap nuts from the 2”...
  • Page 30: Wall Mounted Light Installation

    = 13” PR/PRO SP/SPF = 15” EX/EXL = 11.5” Note: The VistOR PR light head is shown. FIGURE 11: Wall Mount Dimensions GENERAL INFORMATION The shipping cartons contain a light head assembly, arm assembly, a wall bracket assembly with a hospital grade plug attached, hardware kit, and an Installation and Service Manual.
  • Page 31: Figure 12: Wall Mount Installation

    Section 3 Wall Mounted Light Installation Molex Connectors Subassembly Wall Bracket (1) P/N 1000126 Flat Head Cap Screw (1) P/N 0001303 Mounting Hardware (Not included) Cord Retainer Plate (1) Phillips Screws (4) P/N 1002009 P/N 0001056 FIGURE 12: Wall Mount Installation 3 1/2"...
  • Page 32 Section 3 Wall Mounted Light Installation Improper fastening of the wall bracket can cause serious injury and/or property damage. Make certain the installation is capable of supporting a load of at least 50 pounds and an off center moment of 150 ft-lbs. To maintain proper grounding reliability, the ground wire connections with the wall bracket must be kept properly fastened at all times Do not install the extension arm with the light head attached.
  • Page 33: Floor Stand Light Installation

    Section 3 Floor Stand Light Installation Note: The VistOR PR Light is shown. FIGURE 14: Floor Mount Dimensions GENERAL INFORMATION The shipping carton contains a light head assembly, arm assembly, floor base casting, two leg assemblies, a hardware kit, and an Installation and Service Manual. A second carton contains the upright pole assembly.
  • Page 34 Section 3 Floor Stand Light Installation Power Cord Exit location Upright Assembly P/N 1000125 Anti -rotation End Cap (4) Slot P/N 0001206 Adjustment Base Casting Nut (4) P/N 1001170 Flange Nut (4) P/N 0001074 Set Screw P/N 0001128 Caster (2) P/N 0001073 Caster (2) Brake...
  • Page 35: Arm Assembly To Upright Pole Installation

    Section 3 Arm Assembly to Upright Pole Installation Transformer Housing Extension Molex Connectors Mounting Bolts P/N 0001291 FIGURE 16: Arm Assembly to Upright Pole Installation • While securely supporting the extension arm above the upright pole, firmly grasp the cable at the end of the pole.
  • Page 36: Installation Of Light Head To Arm

    FIGURE 17: Installation of Light Head to Arm Installation and repair of this equipment should be performed by qualified persons only. Nuvo, Inc. does not warranty any damage occurring as a result of improper installation. Make sure the light head is installed to the correct arm when assembling multiple lights or combination models.
  • Page 37: Figure 18: Light Head To Arm Electrical Connections

    Section 3 Installation of Light Head to Arm • Insert the yoke/arm connector into the arm until the bolt-holes are aligned. Insert the ¼”-28 x ” bolt as shown being careful not to damage the wires, and secure with the ¼”-28 hex nut. Tighten the nut securely, but do not over tighten.
  • Page 38: Chuttle Track Ceiling Track Mount

    Section 3 Chuttle Track Ceiling Track Mount Installation and repair of this equipment should be performed by qualified personnel only. Nuvo Inc. does not warranty any damage or injuries as a result of improper installation. Trolley Assembly Cable Carrier Junction...
  • Page 39: Chuttle Track Pre-Installation Requirements

    Chuttle Track Pre-Installation Requirements Installation and repair of this equipment should be performed by qualified persons only. Nuvo, Inc. does not warranty any damage or injuries occurring as a result of improper installation. Before installation, check to insure the following conditions are provided and in place.
  • Page 40: Figure 20: Mounting Pallet Perpendicular To Wood Joist

    Section 3 Chuttle Track Pre-Installation Requirements Figure 20: Mounting Pallet Perpendicular Figure 21: Mounting Pallet Parallel to Wood Joist to Wood Joist Figure 22: Mounting Pallet to a Figure 23: Suspended Ceiling Wood Suspended Ceiling Bracing Detail (End View) age 40 of 85 5368 KUHL ROAD ●...
  • Page 41: Disassembly Of The Track Mount Before Installation

    Do not mix the right and left trolley assemblies. The cord bracket within each trolley is positioned to make electrical connection for the right and left light fixtures respectively. Nuvo, Inc. recommends temporarily marking each trolley if needed to make sure these are assembled onto the track in the proper location and orientation.
  • Page 42: Figure 25: Side Panel Removal

    Section 3 Disassembly of the Track Mount before Installation • Take off the two side panels P/N 1001516 by removing the two (2) #10 wood screws at the end of each panel. See figure 25. Side Panels Wood Screws FIGURE 25: Side Panel Removal •...
  • Page 43: Figure 27: Stop Pins Removal

    Section 3 Disassembly of the Track Mount before Installation Keep the side panels and dress covers in a safe place to prevent damage until reassembly. • Remove the two stop pins P/N 1001076 and the #6 flat head screws P/N 0001155 holding the pins in place from the track end with the end plate already removed.
  • Page 44: Track Mount Installation Procedure

    Section 3 Track Mount Installation Procedure Warning: Improper fastening of the track mount can cause serious injury and/or property damage. Insure the installation is capable of supporting a weight of at least 200 pounds and an off center load of 500 ft-lbs. •...
  • Page 45: Figure 29: Ceiling Rod And Roll Pin Installation

    Section 3 Track Mount Installation Procedure • Remove the ¼” by 2” roll pin taped down in the trolley assembly. Insert the pin into the ceiling rod end that has one hole in both walls. The pin should extend the same amount on both sides of the ceiling rod.
  • Page 46: Figure 30: Power Cord Placement

    Section 3 Track Mount Installation Procedure • Ensure the cable carrier chain lays flat within the track. Attach the Molex connector to the mating connector on the trolley assembly. See figures 29 and 30. Ceiling Trolley Assembly Cable Clamp Cable Carrier Cable Carrier should have a Simple Radius FIGURE 30: Power Cord Placement...
  • Page 47: Section 4: Operating Instructions

    Handle Note: VistOR PRO light head shown FIGURE 31: VistOR PR and VistOR PRO Operation age 47 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 48: Vistor Sp And Vistor Spf Operation

    Yoke Sterilizable Handle Main Power Switch FIGURE 32: VistOR SP Spotlight Standard Operation age 48 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 49: Figure 33: Vistor Sp Spotlight Controls

    Section 4 VistOR SP and VistOR SPF Operation • Iris Diaphragm and Dimmer Control Handles: Controls the spot diameter and light intensity for the front lamp. Move the handles left or right as necessary. • Sterilizable Handle: A sterilizable handle is provided on the light head. This handle can be used to position the light head as needed.
  • Page 50: Figure 34: Vistor Spf Spotlight Model Operation

    Fiber Optic Lamp Switch Front Lamp Switch Sterilizable Handle FIGURE 34: VistOR SPF Spotlight Model Operation age 50 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 51: Vistor Ex & Exl Operation

    Yoke Switch Optional Sterilizable Handle (VistOR EXL) FIGURE 35: VistOR EXL Light Operation age 51 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 52: Section 5: Safety Instructions Safety Tips

    After completing a repair of the unit, ensure the unit is in proper working order. Failure to do so could result in personal injury and/or property damage Use only Nuvo approved replacement lamps (refer to “Lamp Replacement Instructions” on Section 6). Failure to do so will affect the operating specifications. Use of a higher wattage lamp may cause a fire hazard, resulting in personal injury and/or property damage.
  • Page 53 Never operate the light with this filter system removed. Failure to do so could result in personal or patient injury. The VistOR PR and VistOR PRO lights come with an internal I.R. filtering system. If the filter appears damaged in any way repair or replace immediately before use on any patient.
  • Page 54: Section 6: Vistor Maintenance

    Unplug the power cord (floor models) before lamps are replaced. Failure to do so could result in personal injury and/or property damage. • For the VistOR PR and VistOR PRO pull off the rear housing cover from the light head as shown in figure 36.
  • Page 55: Figure 36: Vistor Pr/Vistor Pro Lamp Replacement

    Section 6 Lamp Replacement VistOR PR and VistOR PRO Rear Cover Lamp Holder P/N NV1000088 VistOR PR P/N NV1000378 VistOR PRO Head/Yoke Assembly Knob Light Bulb NV0001256 FIGURE 36: VistOR PR/VistOR PRO Lamp Replacement Note: The VistOR PRO Light is shown.
  • Page 56: Initial Lamp Installation/Lamp Replacement Vistor Ex & Exl

    Section 6 Initial Lamp Installation/Lamp Replacement VistOR EX & EXL The VistOR lights operate at high temperatures. Allow the unit or units to cool at least 30 minutes before performing any maintenance. Failure to do so could result in personal injury.
  • Page 57 P/N NV0001257 (VistOR EX) P/N NV0001717 (VistOR OB) FIGURE 37: VistOR EXL Lamp Replacement Note: The VistOR EXL light is shown. age 57 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 58: Lamp Replacement Vistor Sp And Vistor Spf

    • Turn off power and allow the bulbs to cool down approximately 30 minutes. • (Refer to figure 38 for the VistOR SP Spotlight.) Facing the back of the light, lower the housing and keep it in a horizontal position. Using a small Phillips screwdriver, loosen the screw located on the rear grille as shown.
  • Page 59 Section 6 Lamp Replacement VistOR SP and VistOR SPF Lamp Holder Fiber Optic Lamp (Optional) P/N NV0001254 Lever Rear Grille Yoke Sterilizable Handle Fan Motor Socket Front Lamp P/N NV0001255 Housing Lamp Containment Front Lens FIGURE 38: VistOR SP/SPF Lamp...
  • Page 60: Fuse Replacement

    • Replace the blown/broken fuses with Nuvo approved replacement fuses. Use only Nuvo fuses if replacement is necessary. Failure to do so could result in personal injury and/or property damage. To reduce the risk of fire, replace the fuse only with the same type and electrical rating as indicated on the attached table or on the fuse replacement label.
  • Page 61: Figure 39: Fuse Replacement

    P/N NV0001347, 115/120 VAC P/N NV0001462, 230/240 VAC Fuse Replacement Warning Label FIGURE 40: VistOR SPF Fuse Replacement age 61 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 62: Arm Adjustment

    Section 6 Arm Adjustment The arm has been pre-adjusted by the factory. Should further adjustment be necessary, utilize the following procedure. • Remove the cover over the spring tension adjustment slot as shown on figure 41. • Move the arm vertically until the adjustment nut is visible in the slot. •...
  • Page 63: Head/Yoke Adjustment

    Section 6 Head/Yoke Adjustment The head/yoke has been pre-adjusted by the factory. Should further adjustment be necessary, utilize the following procedure. • Find the yoke cap for your specific light as shown in a heavy circle below (see figure 42). •...
  • Page 64: Handle Sterilization

    Assembly Handle Button FIGURE 43: Handle Sterilization Note: The VistOR EXL Exam/Diagnosis light is shown. age 64 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 65: Cleaning Instructions

    Do not use harsh cleaners, solvents, or detergents. Failure to do so could result in equipment damage. The VistOR PRO front lens is supplied with a protective hard coat to resist scratching. Never use abrasive cleaners on front lens. Failure to do so could result in equipment damage.
  • Page 66: Maintenance Schedule

    Note: Maintenance schedules vary for each light depending on usage and operating instructions. An annual inspection of the equipment is recommended at a minimum. Note: Nuvo, Inc. recommends that the maintenance records for this equipment be kept on file at the health care facility.
  • Page 67: Section 7: Troubleshooting General Troubleshooting

    12. Socket is damaged or faulty. 13. No power output from transformer when input power to transformer is measured. VistOR SP Model Only 14. Reconnect wires. 14. Disconnected wires at lamp holder. 15. Turn switch to “On” position. 15. Bulb switch onside of light is in “Off”...
  • Page 68 Section 7 General Troubleshooting (continued) Problem Cause Remedy VistOR SP Models Only 1. Fan blades are obstructed by foreign 1. Remove the foreign object/objects. object. Fan will not turn on 2. Re-connect the fan motor wires. 2. Fan motor wires are disconnected.
  • Page 69: Section 8: Exploded Views And Parts List

    75W, 12VAC (VistOR EX and VistOR EXL) 100W max 12VAC (VistOR PR and VistOR PRO) FIGURE 44: Electric Schematic for VistOR PR/PRO and VistOR EX, 120, 220/230, & 240 VAC age 69 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410...
  • Page 70 Type T4, GY 6.35 Base 75W, 12VAC (VistOR EX/EXL) 100W max 12VAC (VistOR PR/PRO) FIGURE 45: Electric Schematic for VistOR PR/PRO & VistOR EX/EXL, 100 VAC age 70 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410...
  • Page 71: Vistor Sp Electric Schematic

    (set at 152 C) Fan / Motor 24 VAC, 50/60 Hz 17W FIGURE 46: Electric Schematic for VistOR SP, 115/120 VAC & 230 VAC age 71 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410...
  • Page 72: Figure 47: Electric Schematic For Vistor Sp 100 Vac

    (set at 152 C) Fan / Motor 24 VAC, 50/60 Hz 17W FIGURE 47: Electric Schematic for VistOR SP 100 VAC age 72 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 73: Vistor Spf Electric Schematic

    20 VAC Fan / Motor 24 VAC, 50/60 Hz 17 W FIGURE 48: Electric Schematic for VistOR SPF 115/120 VAC age 73 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 74: Figure 49: Electric Schematic For Vistor Spf 220/240 Vac

    20 VAC Fan/Motor 24 VAC, 50/60 Hz 17 W FIGURE 49: Electric Schematic for VistOR SPF 220/240 VAC age 74 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 75: Vistor Pr Exploded View Light Head Assembly

    Section 8 VistOR PR Exploded View Light Head Assembly FIGURE 50: VistOR PR Light Head Assembly age 75 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 76: Vistor Pr Light Head Assembly Part's List

    Section 8 VistOR PR Light Head Assembly Part’s List – P/N 1000087 FIGURE 51: VistOR PR Light Head Assembly, Bill of Materials age 76 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 77: Vistor Pro Exploded View Light Head Assembly

    Section 8 VistOR PRO Exploded View Light Head Assembly FIGURE 52: VistOR PRO Head Assembly age 77 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 78: Vistor Pro Part's List, Light Head Assembly

    Section 8 VistOR PRO Light Head Assembly Part’s List – P/N 1000371 FIGURE 53: VistOR PRO Head Assembly, Bill of Materials age 78 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 79: Vistor Sp Exploded View Light Head Assembly

    Section 8 VistOR SP Exploded View Light Head Assembly FIGURE 54: VistOR SP Spotlight Head Assembly age 79 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 80: Vistor Sp Part's List, Light Head Assembly

    Section 8 VistOR SP Light Head Assembly Part’s List – P/N 1000132 FIGURE 55: VistOR SP Spotlight Head Assembly, Bill of Materials age 80 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 81: Vistor Spf Exploded View Light Head Assembly

    Section 8 VistOR SPF Exploded View Light Head Assembly FIGURE 56: VistOR SPF Head Assembly age 81 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 82: Vistor Spf Part's List, Light Head Assembly

    Section 8 VistOR SPF Light Head Assembly Part’s List – P/N 1000233 FIGURE 57: VistOR SPF Head Assembly, Bill of Materials age 82 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 83: Vistor Ex & Exl Exploded View Light Head Assembly

    Section 8 VistOR EX & EXL Exploded View Light Head Assembly FIGURE 58: EX/EXL Head Assembly age 83 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...
  • Page 84: Figure 59: Vistor Ex & Exl Head Assembly, Bill Of Materials

    VistOR EX & EXL Light Head Assembly Part’s List - P/N 1000381 FIGURE 59: VistOR EX & EXL Head Assembly, Bill of Materials age 84 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410...
  • Page 85 age 85 of 85 5368 KUHL ROAD ● ERIE, PENNSYLVANIA 16510 ● PHONE 814-899-4220 ● FAX 814-899-1410 A member of the Medical Illumination International Group...