DESCRIPTION

1-17.
The HP Model 8 340A Synthesized Sweeper
is a broadband, analog sweep, frequency synthe­
sizer with versatile modulation and output power
control.
1-18.
The 8340A generates synthesized output
frequencies from 10 MHz to 26.5 GHz. Frequency
resolution from 1 Hz at 10 Mhz to 4 Hz at 26.5
GHz is available. Frequency accuracy is a function
of the internal time base accuracy or an external
frequency standard. Harmonically-related signals
are at least 35 dB below the carrier (-35 dBc) and
nonharmonically-related spurious outputs are at
least -5 5 dBc.
1-19.
The 8340A features AM modulation rates
from de to 100 kHz. Pulse modulation circuitry
allows rise and fall times of <25 nanoseconds, an
ON/OFF ratio of> 80 dB, and will pass pulses as
narrow as 100 nanoseconds.
1-20.
Maximum calibrated output power from
the APC 3.5 RF OUTPUT connector is
dBm from 10 MHz to 20 GHz and
from 20 GHz to 26.5 GHz. The lowest output
power available in all bands is -110 dBm. Dis­
played power resolution is 0.05 dB. A Power
Sweep function provides the ability to sweep
power up to 40 dB.
1-21.
The 8340A is capable of generating a true
analog sweep across its entire 10 MHz to 26.5 GHz
frequency range. Additionally, a synthesized
sweep mode allows precision analog sweeps as
narrow as 100 Hz.
1-22.
Additional features include fi v e calibrated
frequency markers and various sweep modes ori­
ented around stored marker positions. The ability
to Save and Recall up to 9 front panel settings also
allows an Alternate Sweep function.
1-23.
The front panel ENTRY DISPLAY shows
the current active function and its appropriate
values. Data entry is quickly accomplished
through· the use of the numeric keyboard, step
keys, or RPG (Rotary Pulse Generator) control.
All front panel functions, except the lines switch,
are programmable through the HP-IB bus.
1-24.
Many serviceability features have also
been incorporated in the 8340A. An internal SELF
TEST is run each time the instrument is turned on
or when the front panel INSTR PRESET key is
pressed. Front panel diagnostic routines are
accessible to aid in fault detection and isolation.
Additional detailed information on troubleshoot­
ing and repair procedures is included in Appendix
A, In Case of Difficulty, in this volume. This infor­
mation is also included in Section VIII, Service, in
Volumes 3 and 4.
1-25.
1-26.
1-27.
a front panel RF OUTPUT connector and a 90 dB
programmable attenuator. Option 001 instru­
ments do not have the 90 dB programmable
attenuator. The lower limit of the output power
range is then -20 dBm.
1-28.
1-29.
OUTPUT connector mounted on the rear panel
instead of the front panel. The 90 dB Programma­
+9
ble Attenuator is installed, thereby allowing the
<
+ 1 dBm
output power level to be set as low as -110 dBm.
<
1-30.
1-31.
PUT connector mounted on the rear panel and the
90 dB Programmable Attenuator is deleted. The
lower limit of the output power level is then -20
dBm.
1-32.
1-32a.
the standard HP 8340A are no longer applicable.
However, HP 8340A Option 006 is still compati­
ble with the 27.8 kHz squarewave modulation
needed to make it operate with HP 8756A or HP
8755 scalar network analyzers.
1-33.
1-33a.
warranted to perform within a relaxed phase noise
specification.
1-34.
1-35.
mount sliding rack mounts on the 8340A. This
allows easier access to the 8340A when mounted
in an equipment rack. It is necessary to remove
t h e i n s t r u m e n t s i d e p a n e l s i n o r d e r t o
Scans by HB9HCA and HB9FSX
General Information

OPTIONS

Option 001, Front Panel RF Output, No
Attenuator
The standard instrument confi g uration has
Option 004, Rear Panel RF Output with
90 dB Attenuator
Option 004 instruments have the RF
Option 005, Rear Panel RF OUTPUT
with No Attenuator
Option 005 instruments have the RF 0 UT­
Option 006, Delete Pulse Modulation
The pulse modulation specifications of

Option 007, Relaxed Phase Noise

Specifications
Instruments equipped with this option are

Option 806, Chassis Mount Slide Kit

This kit includes the necessary hardware to
1-3