Initial Operation

3.

3.1 Initial operation and short instruction

 remove all four transport covers; keep them and re-install them after experiment is done.
 position dilution system inbetween dust or aerosol take-out and measuring instrument
 connect inlet nozzle to sampling probe
 connect measuring instrument to outlet nozzle
 make sure that excess air is drawn away.
No counterpressure may be produced at the excess air outlet.
 toggle switch in position "off"
 connect compressed air supply (4 - 8 bar) (fast coupling NG 8)
 turn pre-pressure control counterclockwise first, until it arrests (lift locking ring)
 switch on dilution stage
 set desired pre-pressure by turning pre-pressure control clockwise
 start measuring instrument and procedure
Escaping aerosols can be toxic and pose a health hazard.
Residual air flows out of the device via the exhaust air outlet.
If hazardous aerosols are used and the exhaust air outlet is open,
these aerosols can get into the environment and harm persons.
► If hazardous aerosols are used: Verify that the exhaust air outlet is
V
V005-06-2020
ERSION
connected to a suitable filter or cleaning system via the exhaust air
tube connector.
O
PERATION
WARNING
M
VKL 10 /100
ANUAL