Leakage Current Limits - GE VIVID E9 Service Manual

Hide thumbs Also See for VIVID E9:
Table of Contents

Advertisement

GE H
EALTHCARE
D
GA091568, R
IRECTION
10-6-2

Leakage Current Limits

ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9.
WARNING
WARNING
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO
VOLTAGE GREATER THAN 30 VOLTS:
1. FOLLOW LOCK OUT/TAG OUT PROCEDURES.
TAG
&
LOCKOUT
2. TURN OFF THE BREAKER.
Date
Signed
3. UNPLUG THE VIVID E9.
4. MAINTAIN CONTROL OF THE POWER PLUG.
5. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE, AS THERE ARE NO TEST POINTS TO
VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON WILL TURN OFF.
BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER
IS TURNED OFF WHEN THE CORD IS STILL PLUGGED INTO THE AC OUTLET.
The following acceptance limits and test conditions are summarized from NFPA 99 and IEC 62353 and
in some cases are lower than that specified by the standards.
In accordance with these standards, fault conditions like Reverse Polarity of the supply mains and Open
Neutral are no longer required for field evaluation of leakage current. Because the main source of
leakage current is the mains supply, there are different acceptance limits depending on the
configuration of the mains (100-120VAC or 230-240VAC).
Compare all safety-test results with safety-test results of previously performed safety tests (e.g.
CAUTION
last year etc). In case of unexplainable abrupt changes of safety-test results consult
experienced authorized service personnel or GE for further analysis.
Table 10-9
Chassis/Enclosure Leakage
Type BF Applied Parts
Type CF Applied Parts
Type CF Applied Parts (sink leakage)
Table 10-10 Leakage Current Limits for System Operation on 230-240 Volt Mains
Chassis/Enclosure Leakage
Type BF Applied Parts
Type CF Applied Parts
Type CF Applied Parts (sink leakage)
10 - 22
5
EVISION
Leakage Current Limits for Operation on 100-120 Volt Mains (US/Canada/Japan)
Leakage Current Test
Leakage Current Test
Section 10-6 - Electrical Safety Tests
System Power
On and Off
On (transmit)
On (transmit)
On and Off
System Power
On
On (transmit)
On (transmit)
On and Off
VIVID E9 S
ERVICE
Grounding/PE
Conductor
Limit mA
Open
0.3
Closed
0.1
Open
0.5
Closed
0.01
Open
0.05
Closed
0.05
Grounding/PE
Conductor
Limit mA
Open and Closed
0.5
Open
0.5
Open
0.05
Closed
0.05
M
ANUAL