Integra DOSE IT P910 Quick Start Manual

Integra DOSE IT P910 Quick Start Manual

Peristaltic pump

Advertisement

Quick Links

Quick Start Guide
This quick start guide is intended to provide a quick overview of your DOSE IT's key
features and to offer basic instructions for getting started. For detailed information,
please refer to the latest operating instructions (OI) that can be found at
biosciences.com
Intended use
This is a general-purpose laboratory instrument. Any use of this instrument in a medical or IVD
setting is the sole responsibility of the user. The programmable peristaltic pump DOSE IT is used to
dispense or pump liquids in a volume range of 0.1 ml to 9999 ml in silicone tubings.
Safety information
Regardless of the listed safety notes, all locally applicable regulations must be observed.
1) DOSE IT may only be used by properly trained personnel in a manner specified by INTEGRA
Biosciences.
2) Observe the hazard warnings on the device. Keep your hands out of rotating wheels inside the
pump head. There is a risk to squash your fingers.
3) Servicing work and repairs may only be performed by INTEGRA Biosciences or an authorized
after-sales service member.
4) Regular servicing of the DOSE IT is recommended.
Getting started
Set up the instrument on a planar surface according to the IQ/OQ document (PN 171273).
Supply voltage: 100
Connect the AC adapter cable to the socket on the rear side of the DOSE IT and plug it to the
electricity mains.
c
d
Switch on the device. The APPLICATIONS window is displayed with three menu options:
Language selection
Press SYSTEM and Language. Select a language and press SAVE. Set other system parameters
to your requirements.
171254_V02_QSG_DOSE_IT_EN.docx
in different languages.
–  
2  
4
0
V
A
, C
5
0
–  
6  
0
a. Operating panel with display and keypad
b. Pump head
a
c. Back-key
d. Main switch on the rear side
b
Switch on/off:
Press the main switch.
• START: To execute the selected program of DISPENSE, PUMP or
CUSTOMIZE type.
• PARAMETER: To define the selected program.
• SYSTEM: To configure general device parameters.
DOSE IT Peristaltic Pump
H
, z
7
0
W
.
www.integra-
1/4

Advertisement

loading

Summary of Contents for Integra DOSE IT P910

  • Page 1 Safety information Regardless of the listed safety notes, all locally applicable regulations must be observed. 1) DOSE IT may only be used by properly trained personnel in a manner specified by INTEGRA Biosciences. 2) Observe the hazard warnings on the device. Keep your hands out of rotating wheels inside the pump head.
  • Page 2 Tubing selection The choice of tubing depends on the requirements for speed and precision of dosing. The larger the inner diameter (ID) of tubing, the faster but less precise is the dispensing. Precision in the table below is expressed as coefficient of variation (CV) and describes the scattering range of the dispenses relative to the mean.
  • Page 3 Equipment Disposal DOSE IT must not be disposed of with unsorted municipal waste. Dispose of DOSE IT in accordance with the regulations in your area governing disposal of devices. Manufacturer INTEGRA Biosciences AG INTEGRA Biosciences Corp. info@integra-biosciences.com CH-7205 Zizers, Switzerland www.integra-biosciences.com...
  • Page 4 171254_V02_QSG_DOSE_IT_EN.docx...

This manual is also suitable for:

171000