CombiFlash Torrent Installation Guide
The system is heavy. Use a two-person lift to prevent injury.
Before making any connections to the Torrent, place it on the bench or in the fume hood
where it will be operated. Temporarily position the Torrent so you can access the back panel
to complete the connections.

2.4 Mounting Optional Modules

If a Solvent Select Valve or Fractionation valve are not being
installed you can skip this section and continue to Section 2.5.
2.4.1 Solvent Select Valve (if purchased)
Place the Solvent Select Valve on the left-hand side of the Torrent
with the valve facing the side of the Torrent.
Use the supplied metal mounting plate and four (4) screws to
secure the Solvent Select Valve to the Torrent.
2.4.2 Fractionation Valve (if purchased)
Place the Fractionation Valve on the right-hand side of the Torrent
with the Valve facing the side of the Torrent.
Use the supplied metal mounting plate and four (4) screws to
secure the Fractionation Valve to the Torrent.
Alternatively, the valve can be connected to a user supplied mast.

2.5 Connect Power

Ensure that the On/Standby switch above the touch screen panel is in the Standby position.
Then, use the supplied IEC power cord to connect the Torrent to mains power.
Mains power must meet the voltage, frequency, and amperage requirements listed on the
serial number label. Input voltage must be within ±10% of the region's nominal line voltage.
In the USA, this is 117 volts ±11.7 volts.
As long as the AC mains power cord is connected to a live outlet, power is inside the unit.
The mains power cord is the disconnect device. Position the Torrent so that the power cord
can be unplugged, or use a power strip where the plug can quickly be removed from the
outlet in the event of an emergency.
Unless power must be removed due to an emergency, always wait at least one minute after
placing the system in Standby before removing the AC mains power cord.

2.6 Connect Interface Cables

Rear panel connections support a variety of devices. Electrical connections to networks or
optional equipment should be completed at this time.
Review the options below and complete as necessary. If any optional modules are not being
installed you can continue to Section 2.7.
WARNING
WARNING
WARNING
CAUTION