Probe Fails; Peripheral Fails; Still Fails; New Unit - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

10-8-3 Probe Fails

NOTE:

10-8-4 Peripheral Fails

10-8-5 Still Fails

10-8-6 New Unit

10-8-7 ECG Fails

Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
When there's too much leakage current ...
Test the probe in another connector to isolate if the fault lies with
the probe or the Ultrasound system. Or Change another probe
to confirm if the fail is caused by console.
Each probe will have some amount of leakage, dependent on its
design. Small variations in probe leakage currents are normal
from probe to probe. Other variations will result from differences
in line voltage and test lead placement. The maximum allowable
leakage current for body surface contact probe differs from
inter-cavity probe. Be sure to enter the correct probe type in the
appropriate space on the check list.
If excessive leakage current is slot dependent, inspect the
system connector for bent pins, poor connections, and ground
continuity.
If the problem remains with the probe, replace the probe.
Tighten all grounds. Ensure star washers are under all ground
studs.
Inspect wiring for bad crimps, poor connections, or damage.
If all else fails, begin isolation by removing the probes, external
peripherals, then the on board ones, one at a time while
monitoring the leakage current measurement.
If the leakage current measurement tests fail on a new
Ultrasound system and if situation can not be corrected, submit
a Safety Failure Report to document the Ultrasound system
problem. Remove Ultrasound system from operation.
Inspect cables for damage or poor connections.
10-31

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid t9

Table of Contents