715 Specifications

Firmware Earlier than V2.0

Input impedance: 2 MΩ (nominal), < 100 pF
Overvoltage protection: 30 V
Voltage drive capability: 1 mA
Overload protection: 125 mA, 250 V fast acting fuse
mA Output: 0 % = 4 mA, 100 % = 20 mA
Temperature Coefficient
Source mode
Simulate mode
Loop Power
1.800.561.8187
715 Calibrator specifications vary based on the version of the instrument. To display the
firmware version for your instrument, start with the 715 off, push and hold , then push
the power button. Find the section heading below for the displayed version and use the
specification tables in that section to test and calibrate the instrument.
Table 15. 715 DC V Input and Output
Range
100 mV
10 V
Table 16. 715 DC mA Input and Output
Range
24 mA
0.005 % of ohms range per °C for temperature ranges – 10 °C to 18 °C and 28 °C to
55 °C
Compliance: 1000 Ω at 20 mA for battery voltage ≥ 6.8 V (700 Ω at 20 mA for battery
voltage 5.8 to 6.8 V)
External loop voltage requirement: 24 V nominal, 30 V maximum, 12 V minimum
24 V ±10 %
www.
Resolution
±(% of Reading + Counts)
0.01 mV
0.001 V
Resolution
±(% of Reading + Counts)
0.001 mA
.com
Process Calibrators
Accuracy,
0.02 % + 2
0.02 % + 2
Accuracy,
0.02 % + 2
information@itm.com
Specifications
13