Enflow Warmer (Model 100 Series) Indicators And Operation - GE enFlow IV System Manual

Fluid and blood warmer
Hide thumbs Also See for enFlow IV:
Table of Contents

Advertisement

Table 1—Controller Display: Normal Operating Model
Activity
Warmer is connected and power is engaged.
Warmer is not connected, but Controller is powered on.
Warmer is connected, but covers are open on Warmer.
Warmer is connected, and covers are either open or
closed on Warmer; however, Cartridge is not in Warmer.
Table 2—Controller Display: Alarm Mode
Activity
Warmer over temperature
Mute button activated
Fault detected
Table 3—Controller Display: Setup Mode
Activity
While powering up the Controller, hold the center button
down until the clock screen is displayed. Then, use the
buttons to set the clock.

enFlow Warmer (Model 100 series) indicators and operation

The Warmer monitors and maintains the infusate temperature at 40 °C ± 2 °C. On the top of the Warmer, there are two status indicator lights (multicolored
LEDs), which reflect the following:
Power - Indicates the power and operational status of the Warmer.
Temperature - Indicates that the infusate temperature is within an acceptable operating range (35 °C to 42 °C).
4400-0024-EN Rev A
Display reads
Temperature and Clock
40 °C
9:00 A
Not Heating
Not Heating
Not Heating
Display reads
Display alternates between Over Temp and
Press Key to Mute
Over Temp Muted
System Fault XX
If a system fault message is on the Controller display,
refer to the system fault section
Appendix B or contact Technical Support.
Display reads
09:00 A
(Indicated by symbol
Sliding covers
ON
OFF
Power cable
Heating surface
- 13 of 40 -
Temperature Status LED
)
TEMP
POWER
Power LED
(Indicated by symbol
)
Display color and function
Identical to Warmer
temperature LED
Yellow
Yellow
Yellow
Display color and function
Identical to Warmer temperature LED
Identical to Warmer temperature LED
Red
High priority alarm
Display color and function
Blue