YOKOGAWA FLEXA FLXA202 Technical Information

YOKOGAWA FLEXA FLXA202 Technical Information

2-wire analyzer hart communication
Hide thumbs Also See for FLEXA FLXA202:

Advertisement

Quick Links

Technical
Information
TI 12A01A02-60E
CONTENTS
What Is HART Communication? ........................................................... 2
Downloading and installing DD files .................................................... 2
Functions Available via HART Communication .................................. 4
Setting Non-Available via HART Communication .............................. 7
Operational Precaution .......................................................................... 7
Note for Functions .................................................................................. 7
(Device revision 2, DD file revision 1)................................................... 8
(Device revision 2, DD file revision 1)................................................. 11
(Device revision 2, DD file revision 1).................................................14
(Device revision 2, DD file revision 1).................................................17
(Device revision 2, DD file revision 1).................................................20
Revision Information .....................................................................................24
FLXA202/FLXA21 2-Wire Analyzer
HART Communication
Before using FieldMate .....................................................................................2
Installing the DD File .........................................................................................3
Connecting Setup Tool .....................................................................................4
Unique Features of HART Communication ....................................................5
Hold status signal: ............................................................................................7
TI 12A01A02-60E
©Copyright Jul. 2013
4th Edition Jun. 2017

Advertisement

Table of Contents
loading

Summary of Contents for YOKOGAWA FLEXA FLXA202

  • Page 1: Table Of Contents

    Technical FLXA202/FLXA21 2-Wire Analyzer HART Communication Information TI 12A01A02-60E CONTENTS What Is HART Communication? ............2 Downloading and installing DD files ............ 2 Before using FieldMate ..................2 Installing the DD File ..................3 Connecting Setup Tool ..................4 Functions Available via HART Communication ........4 Unique Features of HART Communication ............5 Setting Non-Available via HART Communication ......7 Operational Precaution ................7 Setting on the FLXA202/FLXA21 and a setup tool (communication tool): .7 Operation with handling all the parameters of the FLXA202/FLXA21: ..7 Note for Functions ..................
  • Page 2: What Is Hart Communication

    This revision of Device Files can be used with the FLXA202, but the device name will be displayed as FLXA21. Device Files for the FLXA202 are planned to be released in December 2015. TI 12A01A02-60E Jun. 29, 2017-00 No. TI 12A01A02-60E 4th Edition : Jun. 2017(YK) All Rights Reserved. Copyright © 2013, Yokogawa Electric Corporation...
  • Page 3: Installing The Dd File

    Installing the DD File For how to install the DD file, see the respective documentation for the setup tool you use. The following shows how to install the latest DD file in the Yokogawa setup tool, FieldMate, for example. Note that the following descriptions assume installation in FieldMate R2.02.00. If the version is upgraded later, the following descriptions may not match.
  • Page 4: Connecting Setup Tool

    HART Communication Precaution on DD file revision and device revision When using HART communication of a FLXA202/FLXA21, it is necessary to use the DD file that matches the revision of the FLXA202/FLXA21 device used. Make sure to check they match before use, as follows. (1) Checking the FLXA202/FLXA21 device revision Call up the detailed settings screen on the front display panel or access the Zoom display on the setup tool.
  • Page 5: Unique Features Of Hart Communication

    HART Communication l Unique Features of HART Communication (1) Multi-drop mode Up to 15 field devices put in multi-drop mode can be connected in parallel on a single twisted-pair cable. To enable the multi-drop mode, set the device address to a unique number from 1 to 15. Enabling the multi-drop mode fixes the outputs of the corresponding field devices to 4 mA DC and thus disables the 4–20 mA analog output signals as well as burn-out outputs.
  • Page 6 HART Communication (3) Physical write-protect The FLXA202/FLXA21 has two switches for physical write-protection on the back side of the display. Setting two switches to ON position prevents the settings from being changed via HART communication. Write-protect switch The FLXA202/FLXA21 has a write-protect switch for HART communication. Slide switch (for HART only) Reboot switch...
  • Page 7: Setting Non-Available Via Hart Communication

    HART Communication n Setting Non-Available via HART Communication The following FLXA202/FLXA21 settings cannot be done via HART communication: • Changes to matrix information • pH Manual 3Pt • pH Automatic 3Pt • pH Automatic ITP • ORP Manual 2Pt • rH Manual 2Pt • Manual wash (PH, DO) •...
  • Page 8: Dd Menu Structure For Ph Meter Functions (Device Revision 2, Dd File Revision 1)

    HART Communication n DD Menu Structure for pH Meter Functions (Device revision 2, DD file revision 1) Online Sensor1 Zoom Zoon outputs Zoom sensor Sensor wellness Reset wellness data Maintenance Current Time Last calibrated at Calibration due at Zoom device PH module (sensor1) HOUSING ASSY Logbook Read Logbook Add Logbook Execute Calibration pH Calibration pH Manual 1 Point pH Manual 2 Points...
  • Page 9 rH Manual 1 Point Temperature Cal. Hold Temporary output Set temporary output HART Communication Release temporary output Setup Commissioning Sensor setup Measurement setup Temperature settings Temp compensation compensation Process temp. compensation Concentration Calibration settings pH settings Zero and Slope units Limits and timing Buffers Free Program...
  • Page 10 Zero/Slope Impedance Settings Sensor diagnosis settings Sensor diagnosis settings 1/2 Sensor diagnosis settings 2/2 HART Communication Define heat cycle Output setup Simulate mA(Output) Linear mA Table mA Clear AO Table Check AO Values Configure Hold Reset redundant Reset redundant Error configuration Sensor Fault Sensor Warning Housing Fault...
  • Page 11: Dd Menu Structure For Sc Meter Functions (Device Revision 2, Dd File Revision 1)

    X-axis Y-axis Auto Return Adjust contrast Contrast HART Communication MONITOR Display Change Language Change Language Alarm Fault Warning Sensor2 Process Values Basic Setup Device information Loop test Review NOTE In the Device Type Manager (DTM) of FieldMate, the Sensor Wellness menu and Alarm menu are below the Diagnosis menu.
  • Page 12 Zoom device SC module (sensor1) HOUSING ASSY Logbook Read Logbook HART Communication Add Logbook Execute Calibration C.C. Manual SC1 C.C. Manual SC2 C.C. Automatic Air calibration Sample SC1 Sample SC2 Temp. Coef. SC1 Temp. Coef. SC2 Temperature calibration Hold Temporary output Set temporary output Release temporary output Setup...
  • Page 13 Simulate mA(Output) Linear mA Table mA Clear AO Table HART Communication Check AO Values Configure Hold Reset redundant Reset redundant Error configuration Sensor Fault Sensor Warning Housing Fault Housing Warning Logbook configuration Settings logbook Logbook1-1 Logbook1-2 Erase logbook Calculated data setup Advanced setup Settings Settings...
  • Page 14: Dd Menu Structure For Isc Meter Functions (Device Revision 2, Dd File Revision 1)

    MONITOR Display Change Language Change Language Alarm Fault HART Communication Warning Sensor 2 Process Values Basic Setup Device information Loop test Review NOTE In the Device Type Manager (DTM) of FieldMate, the Sensor Wellness menu and Alarm menu are below the Diagnosis menu. n DD Menu Structure for ISC Meter Functions (Device revision 2, DD file revision 1) Online...
  • Page 15 Air calibration Sample SC1 Sample SC2 Temp. Coef. SC1 Temp. Coef. SC2 HART Communication Temperature calibration Hold Temporary output Set temporary output Release temporary output Setup Commissioning Measurement setup Configure sensor Temperature settings Temp compensation compensation Method Temp. comp.1 Temp. comp.2 Concentration Concent.
  • Page 16 Reset redundant Error configuration Sensor Fault Sensor Warning Housing Fault HART Communication Housing Warning Logbook configuration Settings logbook Logbook1-1 Logbook1-2 Erase logbook Calculated data setup Advanced setup Settings Settings Passwords Execute Password Commissioning Password Date/Time Date/Time Communication HART Display Setup Dual display Individual display ISC1...
  • Page 17: Dd Menu Structure For Do Meter Functions (Device Revision 2, Dd File Revision 1)

    HART Communication NOTE In the Device Type Manager (DTM) of FieldMate, the Sensor Wellness menu and Alarm menu are below the Diagnosis menu. n DD Menu Structure for DO Meter Functions (Device revision 2, DD file revision 1) Online Sensor 1 Zoom Zoom outputs Zoom sensor Sensor wellness Reset wellness data Maintenance Current Time Last calibrated at Calibration due at...
  • Page 18 Manual Offset Temperature Cal. Hold Temporary output Set temporary output HART Communication Release temporary output Setup Commissioning Sensor setup Measurement setup Configure Sensor Temperature settings Temp compensation Salinity Compensation Pressure comp.(Measure) Calibration settings Zero/Slope Limits and timing Pressure Comp.(Cal.) Sensor diagnosis settings Define Heat cycle Output setup Simulate...
  • Page 19 Settings Settings Passwords Execute Password HART Communication Commissioning Password Date/Time Date/Time Communication HART Display Setup Dual display Individual display Trend Trend Graph Screen X-axis Y-axis Auto Return Adjust contrast Contrast MONITOR Display Change Language Change Language Alarm Fault Warning Sensor 2 Process Values Basic Setup Device information...
  • Page 20: Dd Menu Structure For Sencom Meter Functions (Device Revision 2, Dd File Revision 1)

    HART Communication n DD Menu Structure for SENCOM Meter Functions (Device revision 2, DD file revision 1) Menu(Online) Online Sensor1 Zoom Zoom outputs Zoom sensor Sensor wellness Rest wellness data SENCOM Sensor status Last date of Sterilization Last date of High temp.1 Last date of High temp.2 Maintenance Current Time Last calibrated at Calibration due at Zoom device PH module(sensor1)
  • Page 21 pH Sample ORP Calibration ORP Manual 1 Point rH Calibration rH Manual 1 Point HART Communication Temperature Cal. Hold Temporary output Set temporary output Release temporary output Setup Commissioning Sensor setup Sensor remove Measurement setup Temperature settings Temp compensation compensation Process temp.
  • Page 22 Zero/Slope Validate rH/ORP cal. coefficient rH settings Limits and timing Zero/Slope HART Communication Validate rH/ORP cal. coefficient Impedance Settings Sensor diagnosis settings Sensor diagnosis settings 1/2 Sensor diagnosis settings 2/2 Define heat cycle Define SENCOM status Output setup Simulate mA(Output) Table mA Clear AO Table Check AO Values...
  • Page 23 HART Display Setup Dual display Individual display HART Communication Trend Trend Graph Screen X-axis Y-axis Auto Return Adjust contrast Contrast MONITOR Display Change Language Change Language Alarm Fault Warning Sensor2 Process Values Basic Setup Device information Loop test Review NOTE In the Device Type Manager (DTM) of FieldMate, the Sensor Wellness menu and Alarm menu are below the Diagnosis menu.
  • Page 24: Revision Information

    Addition of SENCOM July 2013/1st Edition Newly published (based on Appendix 4 of IM 12A01A02-01E 3rd Edition) n If you want have more information about Yokogawa products, you can visit Yokogawa’s homepage at the following web site. Homepage: http://www.yokogawa.com/an/ n Written by Analytical Products Marketing Dept.