Setting Up Your ResMed Connectivity Module (RCM)™
Important Points:
The RCM™ can only be connected to one device at a time. It is best used with your
main device that you use most often if you have more than one device.
The RCM™ is ideally placed 12" away from your Stellar150™ or other electrical
equipment and 1m (3'3") away from mobile communication devices such as a cell
phone.
Make sure the RCM™ is placed more than 2cm (0.8") away from your body during
operation.
©2020 Ontario Ventilator Equipment Pool. All rights reserved.
with the Stellar150™
Operating State
Power - Green: ON: power is connected and on
OFF: power is off
Vent. Input - Blue: ON: connected to the Stellar150™
OFF: disconnected
BLINKING: connecting to Stellar150™
Note: The Blue Vent. Input led will be OFF when the
Stellar150™ device is off but will come back on when the
Stellar150™ unit is turned back on to start therapy
Signal – Blue: ON: indicates connectivity to the cellular
network and signal strength
OFF: no cellular network detected
Note: The RCM™ needs to detect a cellular network in
order to transmit daily data. Contact the VEP if your
device is not detecting a network in your area.
Error Indicator: If ON, unplug and plug the RCM™ back in.
If this does not remove the error call the VEP for
assistance in troubleshooting or replacement if needed