Chapter 3, Operating Overview
Using Remote Control

Using an ASCII Terminal

86
S:\HP892XX\USRGUIDE\manual\opoverv.fb
Connecting an ASCII terminal to the serial port allows you to remotely operate
the Test Set by entering characters that represent each front-panel control.
Before you can use this feature, you must first set the required serial port settings
in the I/O CONFIGURE screen, and make any hardware connections.
The Serial Port connections are described in
.
Knob Descriptions"

To Configure for Serial Port Operation

1. Access the I/O CONFIGURE screen.
2. Set the Serial In field to Inst.
3. Set the IBASIC Echo field to On.
4. Set the Inst Echo field to On.
5. Set the remaining serial communications fields according to your terminal/computer's
serial communication requirements. These fields include:
Serial Baud
Parity
Data Length
Stop Length
Rcv Pace
Xmt Pace
6. The Test Set now responds to the equivalent characters sent to it by the terminal/com-
puter.

Equivalent Front-Panel Control Characters

lists the terminal/computer keystrokes that equate to front-
Table 6 on page 87
panel controls. Each sequence must be preceded by the Escape key.
For example, to remotely access the CONFIGURE screen, press the Esc key, then
press the C key on your terminal/computer. Be sure to use upper-case C for this
example.
chapter 24, "Connector, Key, and