  • Page 1 ViZion DR+ 1417CK Calibration Guide...
  • Page 2: Table Of Contents

    ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X Contents Calibrating ViZion DR+ (1417CK) ......................2 Preparing for Calibration ......................... 2 Recommended Panel Configurations ..................2 Powering on the Panel ......................2 Stopping the Maven Process ....................3 iDemo ............................. 3 Connecting to iDemo .......................
  • Page 3: Calibrating Vizion Dr+ (1417Ck)

    Calibrating ViZion DR+ ViZion DR+ 1417CK Wireless Calibration Guide v1.0 (1417CK) Ultra UAI v4.1.0.X This guide provides detailed instructions on how to run a full gain and defect calibration for your ViZion DR+ 14x17CK wireless panel once the installation process has been tested and verified.
  • Page 4: Stopping The Maven Process

    ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 3. Slide the two battery lock levers until it 1.3 Stopping the Maven Process clicks into place. 1. Exit the ViZion DR+ acquisition software. 2. Stop the Maven process if it is running.
  • Page 5: Configuring Idemo

    ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 2. Click Configuration and change FTP 2.2 Configuring iDemo port to 21000 and click Apply. 1. In the Correction section, configure the following settings. 3. Click Start and minimize the FTP Server window (Do not close).
  • Page 6: Calibrating Vizion Dr

    ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 3. Click General Settings on the left 3 Calibrating ViZion DR+ menu, set the following general configuration settings. 3.1 Environment Setup 1. Align the panel with the X-ray tube. Trigger = Inner Trigger 2.
  • Page 7 ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 2. Click Create. 7. After image transfer completes and the status “Got Image” is displayed, verify 3. Select the Gain tab. that the Center Average value is between 5900-6100. Figure 14: Gain Settings 4.
  • Page 8 ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 9. When center average value is within the 11. Click Create Gain Correct File. expected range of 5900-6100, click Select (Figure 17) to add that value to the Selected Points list (Figure 18).
  • Page 9: Performing Panel Defect Calibration

    ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 4. Configure the dose as close to the 3.3 Performing Panel Defect requested kV as possible, trying to Calibration adjust mAs to get within ± 100 of the NOTE: This step is optional. After...
  • Page 10 ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 6. Click OK after you have read the advance to the next number and a pop- expected dose information message. A up for the next frame may display. message in red letters “Waiting for...
  • Page 11: Storing New Calibration Files

    ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 10. Once you successfully complete a defect frame, the next “Current” # should increase to the next # in the sequence automatically (increase it manually by clicking on the dropdown...
  • Page 12 ViZion DR+ 1417CK Wireless Calibration Guide v1.0 Ultra UAI v4.1.0.X 6. Click Write RAM to save changes. Figure 28: HW Correction 5. If you also performed a new Defect Figure 29: HW Correction Calibration, click Download in line with 7. Click Exit to exit the Config Screen.