Graphical Display; Turn On The Analyzer; Start The Analyzer Setup - Hach 5500sc Operations

Ammonia/monochloramine
Table of Contents

Advertisement

Available languages

Available languages

Graphical display

The graph shows measurements for a maximum of four parameter at once. The graph supplies easy
monitoring of trends and shows changes in the process.
1. From the main measurement screen, push the LEFT arrow to show the graphical display.
Note: Push the UP or DOWN key to show the graph for the previous or next measurement parameter in
sequence.
2. Push home to change the graph settings.
3. Select an option.
Option
MEASUREMENT VALUE Set the measurement value for the selected channel. Select between AUTO
DATE & TIME RANGE
Startup

Turn on the analyzer

1. Open the top door.
2. Pull the analytics panel open. A magnetic latch holds the panel closed.
3. Turn on the power switch on the main circuit board (refer to
4. Close the analytics panel.

Start the analyzer setup

When the analyzer power is set to on for the first time or after the configuration settings were reset to
the default values:
1. Select the applicable LANGUAGE.
2. Select the DATE FORMAT.
3. Set the DATE and TIME.
4. When prompted, confirm with YES to start the automatic analyzer setup.
Note: The analyzer stays in initialization mode until the configuration is complete.
5. Select the channel.
6. Push enter to confirm.
7. When prompted, make sure that the reagent bottles are full and confirm with YES.
8. When prompted, make sure that the cleaning solution bottle is full and confirm with YES.
9. When prompted, make sure that the standard bottles are full and confirm with YES.
10. Push enter to start the reagent priming process.
Note: The analyzer setup does not configure relays, outputs, network cards, calculations or calibration parameters.
Refer to
Calibration
on page 19 for configuration information for calibration parameters. Refer to
outputs
on page 12 for configuration information for relays, outputs or network cards. Refer to
calculation
on page 12 to setup calculations.
8 English
Description
SCALE and MANUALLY SCALE. Enter the minimum and maximum ppb value in
the MANUALLY SCALE menu.
Select the date and time range to show on the graph: last day, last 48 hours, last
week or last month.
Figure 1
on page 5).
Configure the
Set up the

Advertisement

Table of Contents
loading

Table of Contents