Batteries and charging

Procedures for Charging.

24V 10A charger
The 24V 10A charger has one
bi-coloured orange/green indicator
to show the charger operation
(Fig. 29).
1 Plug into the three pin charging
point under the fornt of the
control box (Fig. 28).
2 Connect the charger to the
mains supply by means of the
mains plug and switch on.
3 At the start of charging the
charger will undergo a self test
procedure, this is indicated by the
indicator flashing, alternating
orange/green (Fig. 29).
4 When the test is complete and
there have been no faults found,
charging will commence. This is
indicated by a flashing orange
only light.
5 On return of 90% or the bulk of
the charge a steady orange light
will be visible.
6 When a steady green light is
visible the battery is fully charged.
7 An automatic switch to a float
voltage will occur after a further
period of time, and the charger
should be left connected to the
wheelchair until it is required for
use.
8 No light indicator will show
unless the charging plugs are
connected.
9 Always switch off at the mains
before disconnecting the
batteries.
Control box
Charger
socket
Fig. 28
Orange/green
warning light
Fig. 29
27
On/Off switch
on rear case
Issue 4