Honeywell VM242A BasicMes-2 Operating Instructions Manual page 16

Handheld flow measuring computer
Table of Contents

Advertisement

VM242A BASICMES-2
Technical specifications
Basic device specification
Medium
Medium temperature
Ambient temperature
Storage temperature
Static (burst) pressure
Differential pressure
Differential pressure
resolution
Accuracy
Scan rate
Interface
Weight
Dimensions
NOTE: Storage temperature below 2°C (36°F) only when
device and pressure hose assembly are drained
Units and Display
Pressure units
mbar, bar, hPa, kPa, mH
Temperature units °C, °F
Flow display
up to 5mbar: no display
0...1,000l/h: resolution 1l/h
1...100m
100...1,000m
1,000...10,000m
k
-value input
k
< 1 in increments of 0.001
v
v
k
1...99.99 in increments of 0.01
v
k
100...999.9 in increments of 0.1
v
k
1,000...10,000 in increments of 1
v
Density correction
0.10...5.00kg/l in increments of 0.01kg/l
factor
Internal Temperature Measurement
Measuring range
-20...60°C (-4...140°F)
Accuracy
< ±1K
Resolution
0.1°C
EN0H-2323GE25 R0111
Water or glycol-water mixture,
glycol content max. 50%. Quality
to VDI 2035
-20...120°C (-4...248°F)
5...40°C (41...104°F)
-20...60°C (-4...140°F); see note
below
max. 32bar (464psi)
max. ± 17bar (247psi)
up to 1bar: 0.1mbar
above 1bar: 1mbar
<3% of measured value
in range of ±10mbar better than
±0.3mbar at stable conditions in
ambient temperature range
Normal: 1Hz, average value of four
scans (time slot four seconds)
Fast: 4Hz, no averaging
USB, HP-IR for communication
with pocket printer
2.5kg including accessories and
carry case
470 x 370 x 110mm
(length x width x height)
O, inH
2
2
3
3
/h: resolution 0.01m
/h
3
/h: resolution 0.1m
3
/h: resolution 1m
External Temperature Measurement
Measuring range
Accuracy
Resolution
Power
Power supply
Power consumption Typically 70mA with display brightness
Languages
Standard languages English, Dutch, French, German,
For other countries a flash upgrade can be downloaded from
the DocuServer.
Eastern Europe
Northern/Southern
Europe
Scope of Delivery
• VM242A BasicMes-2 handheld measuring computer with
four rechargeable batteries and lanyard
• Set of hoses with bypass assembly and Honeywell
O, psi
TM
SafeCon
quick connections
• Battery charger
• USB cable
• Two adapters Honeywell SafeCon
connection of BasicMes-2 to Honeywell valves with old style
3
/h
PT valves)
3
/h
• Two adapters Rectus 21 onto 3/4" internal thread
• Carry case with foam insert
• User software on CD-ROM
• Instruction manual with pocket guide
Ordering Information
OS-No.
Description
VM242A0101 Honeywell VM242A BasicMes-2 handheld flow
measuring computer
2
-20...300°C (-4...572°F)
±2K from 0°C to 133°C
otherwise 1.5% of actual value,
according to EN50379-2
0.1°C
Four rechargeable batteries type AA,
mains charging adapter supplied with
device
of 60% (default)
max. 120mA with 100% brightness
50µA for real time clock during power
down and log mode
Italian and Spanish
English, Czech, Hungarian, Polish and
Slovakian
English, Danish, Swedish and Turkish
TM
onto Rectus 21 (for
Honeywell Subject to change

Hide quick links:

Advertisement

Table of Contents
loading