  • Page 1 OPERATION MAUNAL IL3 series Model : IL3-15 / IL3-25 Manual No. : 220412L002 Version : 0.0...
  • Page 2 Before using this product, read this entire Operator's Manual carefully. Users should follow all of the Operational Guidelines contained in this Manual and take all necessary safety precautions while using this product. Failure to follow these guidelines could result in potentially irreparable bodily harm and/or property damage.
  • Page 3 Quality Management System Jeio Tech Co, Ltd. is dedicated to providing world-best product quality customer satisfaction. ensure maintain this commitment have developed implemented a total quality program, which conforms to the requirements according to DIN EN ISO 9001:2000 for the design, development, production, sales and servicing of biotechnology, environmental chemical engineering related products, and reliable measuring equipment for...
  • Page 4: Table Of Contents

    Table of Content 1.0 Safety ........................................1 1.1 How to use the Manual ..........................2 1.1.1 Introduction ....................................2 1.1.2 Chapter summary ................................... 2 1.2 Responsibility ..............................2 1.2.2 Responsibility of the users ..............................3 1.2.3 Preventative Maintenance ..............................4 1.3 Symbols used in this Manual ........................
  • Page 5 4.6.2 Temp Calibration ................................. 29 4.6.3 Temp Deviation ..................................31 4.7 Warning and Fault Messages ........................32 4.7.1 Warning ....................................32 4.7.2 Fault ......................................33 4.7.3 Other notifications ................................34 4.8 Safety ................................35 4.8.1 Over Temperature Protection –Electronic temperature limiter ................35 4.8.2 Current Protection ................................
  • Page 7: Safety

    1.0 Safety 1 Manual NO. 220412L002 Version : 0.0...
  • Page 8: How To Use The Manual

    1.1 How to use the Manual 1.1.1 Introduction This manual is intended for individuals requiring information about the use of product. Use this manual as a guide and reference for installing, operating, and maintaining your Jeio Tech product. The purpose is to assist you in applying efficient, proven techniques that enhance equipment productivity This manual covers only light corrective maintenance.
  • Page 9: Responsibility Of The Users

    1.2.1 General Liability In any case, safety is the first priority. The owner of the machine, user, and/or maintenance personnel must be fully aware of the safety which is the most important part of our everyday business. If you are concern about the loss in efficiency, remember that accidental impact is always negative for the effectiveness of covering human resources and property.
  • Page 10: Preventative Maintenance

    • When not in use, turn off the power switch or cut off the power to the instrument by disconnecting the power plug. 1.2.3 Preventative Maintenance Proper maintenance is essential for safety. The maintenance personnel must be aware of the importance of safety in maintenance of the equipment for effectiveness of the unit. Before removing or adjusting parts of the device, shut off all the electrical supply, remove accessories.
  • Page 11: Safety Symbols

    1.3.2 Safety symbols Mandatory Read Manual. Wear gloves. Wear eye Wear a face mask. protection. Prohibition No high frequency No direct sunlight. noise. No corrosive fluids or No water. cleaners. 5 Manual NO. 220412L002 Version : 0.0...
  • Page 12 Warning Foot crush. Hand crush or pinch. Do not take the device Lifting hazard. apart deliberately. Electrical shock. Flammable or fire could be caused. Sharp points. Hot surface. Safety Alert Symbol Crush hazard. General caution. 6 ©2016. All Rights Reserved. Jeio Tech Co., Ltd...
  • Page 13: Non-Safety Symbols

    1.3.3 Non-safety symbols The following graphic representations are intended to convey a message without words or to bring your attention to important information about the use of the Product or a feature. Permissible ambient Altitude temperature Relative humidity Earth Ground 1.4 Where to Locate Safety Labels on product.
  • Page 14: Warning Statements

    1.5.1 Warning statements Indicates a hazardous situation which, if not avoided, could result in death or serious injury Observe all warning labels. DO NOT remove warning labels. Check the voltage, phase and capacity of the power supply and connect properly. Power supply must be properly grounded.
  • Page 15: Caution Statement

    DO NOT operate damaged or leaking unit. DO NOT disassemble, fix or change the Product other than for those items described in this operating manual. In case of inner electronic components service, only the person who is qualified can handle the problem.
  • Page 16 Do not inject any flammable objects, and conductive objects outside of product hole or vent. Wear safety gloves and be careful burn from the residual heat from the product. Do not install near other laboratory instrument, electrical devices, or power terminal. In case of inner electronic components service, only the person who is qualified can handle the problem.
  • Page 17: Functional Description

    2.0 Functional Description 11 Manual NO. 220412L002 Version : 0.0...
  • Page 18: Introductions

    2.1 Introductions This equipment is a multi-purpose low temperature incubator for biotechnology, pharmacy, medical science, chemistry, and biology. It has applied CLS (Custom Logical Safe) – Control System that Jeio Tech originally developed for the convenience in use and the s 2.2 Features 1) CLS-Control system means “Control system which has logical safety device specialized for...
  • Page 19: Structure

    2.3 Structure [ IL3-15 ] [ IL3-25 ] 13 Manual NO. 220412L002 Version : 0.0...
  • Page 20 (1) Temperature Controller This has a micro processor (CPU) which has a digital PID auto tuning function. It also has the highest level of safety control system such as a temperature compensational function for temperature sensor and a heating volume controller. (2) Door For using a silicone heater near door, we removed dewdrops on the surface of the door.
  • Page 21: Installation

    3.0 Installation 15 Manual NO. 220412L002 Version : 0.0...
  • Page 22: Components

    3.1 Components After unpacking, please check the contents to ensure you have received all of the following unit components. Also, check the identification plate on the side of the unit to make sure you received the model number your ordered. If you didn’t receive one or more of the components or if the model is incorrect, contact your local Jeio Tech office, or the distributor from which the unit was purchased.
  • Page 23: Environmental Setting

    3.2 Environmental setting The unit can be operated properly under the following environmental conditions for a long time running without any problem. No direct sunlight. Keep product away from high frequency noise produced by equipment and/or machinery, such as electrical distribution panels, welders, induction heating mechanism, and large amounts of SCR (Silicon- controlled rectifiers).
  • Page 24: Power Supply

    Temperature and humidity (5) Please avoided from heat sources and direct sunlight, installed condition: 18°C ~ 25°C, below 80%RH condition (6) The product should be avoided to use near organic solvents such as Acetone, Methylene Chloride, etc. also do not use near strong and high frequency noise place. (7) Do not allow any moisture, organic solvents, dust, or corrosive gases to enter the control box of the unit.
  • Page 25: Operation

    • Check the voltage, Phase, current capacity exactly to connect the power properly. • Inappropriate connection of power can occur damage of the unit and serious injury to the user. Electrical Shock • Do not touch or deal with the electric parts and cord with wet hand. Hazard.
  • Page 26: Operation

    4.0 Operation 20 ©2016. All Rights Reserved. Jeio Tech Co., Ltd...
  • Page 27: Function Of Control Panel

    4.1 Function of control panel The control panel is made up with VFD, Dial Knob, and Membrane button. Status Display Display the status of the unit. • Run: The unit is running • A-tune: The unit is on the Auto-tuning. •...
  • Page 28: Control Structure

    * Wheel the Knob to select the configuration. Push the Knob to go to the selected menu. * Push the Knob for long time to bring the saved set temperature values. Control structure The main structure of the unit is as below. Press the dial knob for one time, and set the temperature value.
  • Page 29: Basic Operation

    Basic operation 4.3.1 Operation Turn on the switch located in right below of control panel. VFD will be on with “JEIOTECH” and Main screen will be displayed. 4.3.2 Temperature control 4.3.2.1 Start to temperature control When you push Start/Stop button to control the temperature when the status of Run refer to Starting temperature confrol.4.3.1 4.3.2.2 Stop to temperature control...
  • Page 30: Controller Lock

    If no input in the knob for 20seconds, the setting will be canceled. (3) Push the Knob to save the adjusted setting. ※ In case of Timer setting, the procedure is the same as above. ※ In case of Menu configuration, Push Menu button while selecting the setting and change the setting as above.
  • Page 31: Timer Off Setting

    (2) Wheel the Knob to select the timer type (Wait on, Wait off) and press the knob to enter the setting menu. [Wait On Timer] [Wait Off Timer] (3) Wheel the Knob to the left to select TIMER SET and press the knob. The hour value will be blinking.
  • Page 32: System Configuration

    Fixed Temp Set You can register 3 sets of fixed temp. set. You can bring the setting pressing the Knob for 3seconds. Unit Temp. You can choose the temp. measurement.(C, F) Sound You can on and off the beep sound of the system. Display You can choose the status of the display when you wheel the knob while the unit is running.
  • Page 33: Auto Tune

    4.6.1 Auto Tune The function performs updating PID parameters for correct temperature control. It makes the unit can perform the best temperature control in main using temperature for user. If the unit can’t control the temperature at first when the user uses the unit, use Auto Tuning function to tune the temperature control parameters.
  • Page 34 PV on performing the Auto tuning Indicates Auto tuning procedure. You can cancel the autotuning pressing Knob or STOP button (4) You can see the completion message when the tuning is done. Press the knob to save the tuning value. (5) Press MENU button >...
  • Page 35: Temp Calibration

    4.6.2 Temp Calibration You can calibrate the temperature sensor using this function. The unit can tune the temperature with 3 measure point and user’s reference measuring equipment. After calibration, the unit recognizes the temperature in the chamber as the one read by user’s reference measuring equipment. Enter the menu following below procedure and the unit can perform temperature calibration with 3points (CALI1, CALI2, CALI3).
  • Page 36 (1) Select MENU > SYSTEM CONFI > CALIB TEMP to enter below menu. (2) Choose CALI 1~3 using control knob and set the temperature for calibration. [In case of setting CALI2 on 50C] (3) After actual temperature inside of the chamber reaches setting value, press the knob to enter the temperature value which user wants to have calibration.
  • Page 37: Temp Deviation

    (6) Rotate the control knob to the end and select ESC (or no input for 20seconds) to go back to main display.  If the error occurs such as exceed the range of over temperature during temperature control, the control stops. ...
  • Page 38: Warning And Fault Messages

    4.7 Warning and Fault Messages The unit alarms user when detecting abnormal state. It can draw your attention with visual alerts. The abnormal condition can be divided depending on serious one (Warning) and normal one (fault). 4.7.1 Warning Warning: Audiovisual alarm is on but the unit is in operation. User can stop the audio alarm by pressing knob and Knob but visual alarm is on until the problem disappears.
  • Page 39: Fault

    4.7.2 Fault This chart show the fault classified which is made product forced stop. Press the Knob, then the visual warring will be release. The product will be stopped the operation for protect themselves. Over Activated the electronic temperature limiter. In case of over Temperature temperature in chamber, the main power will be cut off due to the temperature limiter for the protect sample and product.
  • Page 40: Other Notifications

    4.7.3 Other notifications When user press Knob butten, the screen will be changed from “Notice” to “Main Display”. Auto Start Notice : to show that the product power is cut off by forcibly, and supplied again When user set the “A.RUN” to”ON” in configuration, the display showen “A.RUN”...
  • Page 41: Safety

    4.8 Safety 4.8.1 Over Temperature Protection –Electronic temperature limiter The electronic temperature limiter is basically equipped (Based on the temperature value of the NTC temperature sensor). The limiter will be stopped the switch off, In case of the chamber temperature become a higher or lower from limiter set value.
  • Page 42: Current Protection

    User can check the parameter setting value 9 using (Config button), and also change the parameter setting value (by Up/Down button) Lower limit: -10 ~ upper limit (by Celsius) Upper limit: lower limit to 120 degrees (Celsius criteria) The screen will be returned to the main screen, If the product is standby condition for 5 sec. If the temperature limiter shown value is different with actual value (measured calibration temperature device), Please corrected in the following way;...
  • Page 43: Maintenance

    5.0 Maintenance 37 Manual NO. 220412L002 Version : 0.0...
  • Page 44: Inspection Cycle

    5.1 Inspection Cycle INSPECTION TIME CLASSIFICATION FRAME DAILY WEEKLY ● Inspect power cord connection at unit and receptacle. ● Inspect power cord for wear, cracks or cuts. Cleanliness inner/ Outer of chamber ● Cleanliness of Accessories ● Inspect the operating the start/ stop button, dial ●...
  • Page 45: Move

    Move (1) When moving the unit, be sure to unplug the power cord from the wall outlet. (2) Remove all the stuff inside of the unit. (3) Pack the unit and accessories with proper packing (original packaging or any other proper packing) to protect the product being damaged during the movement of the unit.
  • Page 46: Troubleshooting

    6.0 Troubleshooting 40 ©2016. All Rights Reserved. Jeio Tech Co., Ltd...
  • Page 47: Malfunction Check List

    6.1 Malfunction Check list Symptoms Check list Measures Check the ID plate which is attached to the unit and properly connect the power Check for improper power supply type (voltage, phase, capacity) supplied to the outlet. Power failure, Check the panel board breaker and Check the operation of the restore it to its original state.
  • Page 48 Remove all the devices connected to the breaker and connect them one by Too many plugs are one to find out the cause. Do not Leakage of the building If connected. exceed the capacity of the earth circuit breaker continues short- leakage breaker current.
  • Page 49: Jeiotech Software

    7.0 Jeiotech Software 43 Manual NO. 220412L002 Version : 0.0...
  • Page 50: Installing Monitoring Program

    7.1 Installing Monitoring Program When you insert the installation CD into the CD-ROM drive, the installation program starts automatically. Click the "Next" button to switch to the License agreement screen. (2) Before installing JeioTechSoft, review the contents of the license and click the "I agree" button.
  • Page 51 When the program installation is completed, JeioTechSoft icon will be created on the desktop as shown below. Run by JeioTechSoft 2.0.0.5 Check the list box or double click on the desktop icon to run the program. • During JeioTechSoft program execution, input values of temperature, timer, and pump circulation amount can be directly by numeric keypad of computer keyboard or by ↑...
  • Page 52: Appendix

    8.0 Appendix 46 ©2016. All Rights Reserved. Jeio Tech Co., Ltd...
  • Page 53: Technical Specifications

    Technical Specifications Model IL3-15 IL3-25 IL3-15A IL3-25A 150/5.3 242/8.5 156/5.5 254/8.97 Chamber volume (L/cu ft) Temperature 0 to 80/32 to 176 4 to 70/39.2 to 158 Range (℃/ ℉ 0.1/0.18 0.2/0.36 Fluctuation at 37℃ (±℃ ℉ 0.8/1.44 1.0 /0.18 0.7/1.26 1.0 /0.18...
  • Page 54: Disposal Of Instrument

    8.2 Disposal of instrument Before disposing of the device or its components, observe the following precautions. (1) This equipment is enough to potentially exposed to biological, chemical, or radioactive material, so the person responsible for disposing of the equipment and the waste disposer need to check essentially that the equipment should be cleaned for the purpose of protecting the environment Please refer to the person in charge of disposal of the...
  • Page 55: Warranty

    8.3 Warranty 9.3.1 Terms of Warranty Service (1) Customer can get free warranty service for 2 years limited warranty from the date of purchase when the machine is broken while operating (2) When user order to repair, please check the below details. 9.3.2 Warranty exception Customer can’t get free warranty service in case of as below.
  • Page 56: Technical Assistance

    8.4 Technical Assistance International Sales Head Office (Korea)  #1005, Byucksan Digital Valley 6-cha, 219, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea 153-704 Tel: +82 2 2627 3816 E-mail: overseas@jeiotech.com FAX: +82 2 3143 1824 The Americas (U.S.A. Branch)  19 Alexander RD Unit #7, Billerica MA 01821, U.S.A. Tel: +1 781 376 0700 E-mail: info@jeiotech.com...

This manual is also suitable for:

Il3-25

Table of Contents