Errors And Warnings - Hanna Instruments 96711C Instruction Manual

Free and total chlorine ism
Table of Contents

Advertisement

Distributed by Applied Membranes, Inc.
http://www.appliedmembranes.com
DISPLAY ELEMENTS DESCRIPTION
5
6
4
3
2
1
1) The measuring scheme (lamp, cuvette, detector), appears during different phases of zero or
reading measurement
2) Error messages and warnings
3) The battery icon indicates the charge state of the battery
4) The hourglass appears when an internal check is in progress
5) Status messages
6) The chronometer appears when the reaction timer is running
7) The month, day and date icons appear when a date is displayed
8) Four digit main display
9) Measuring units
10) Four digit secondary display
10
7
8
9
10
Distributed by Applied Membranes, Inc.
http://www.appliedmembranes.com

ERRORS AND WARNINGS

The instrument shows clear messages when erroneous condition appears. Messages are also displayed
when the obtained values are outside expected range. The beeper is playing a beep on errors.
a) on zero reading
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
No Light: The instrument cannot adjust the light level. Please
check that the sample does not contain any debris.
b) on sample reading
Inverted cuvettes: The sample and the zero cuvette are inverted.
Zero: A zero reading was not taken. Follow the instructions of
the measurement procedure for zeroing the meter.