Self Test - Agilent Technologies J-BERT N4903 User Manual

High-performance serial bert
Table of Contents

Advertisement

9
Customizing the Instrument

Self Test

Self Test - Concepts
N O T E
N O T E
Self Test - Procedures
426
The self test checks specific hardware components for basic
functionality. At initialization, the following self tests are automatically
run:
Pattern Generator Power Up Test
Error Detector Power Up Test
The current system state is saved and the instrument is re-initialized before
performing the selected tests. The saved system state is restored upon completion
of the self test.
When Self Test is selected from the Utility menu, the Self Test Options dialog
box opens. In this dialog box, you can select additional self tests that
will run when you click Run Test.
During accumulation or while auto-search functions are in progress, Self Test is
not available.
The pass/fail results of the self test are displayed in the Self Test
Options dialog box. In case of failed tests, you can view detailed error
messages in a separate window.
To run a self test:
1 Select Self Test from the Utility menu.
This opens the Self Test Options dialog box.
2 Select the desired tests and click the Run Test button.
3 After the self test has finished, the results are displayed next to each
selected test.
Agilent J-BERT N4903 High-Performance Serial BERT

Advertisement

Table of Contents
loading

Table of Contents