Siemens SITRANS F MAG 8000 Standard Operating Instructions Manual page 86

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F MAG 8000 Standard:
Table of Contents

Advertisement

Troubleshooting/FAQs
9.1 Fault codes
Fault
Name/text
codes
4
Database
checksum
5
Low Power
Alarm
6
Flow Overload Flow rate has exceeded
7
PulseA over‐
load
8
PulseB over‐
load
9
Consumption
Interval
86
Description
Corrupted data in ee‐
prom detected by
checksum test made
during power-up
Battery capacity is be‐
low preset threshold
(default 10%).
Power up (temporary
warning – will disappear
after 4 hours)
125% of Q3
Duty cycle of output A
has exceeded maxi‐
mum possible of 50
Duty cycle of output B
has exceed maximum
possible of 50
Accumulated volume
on totalizer 1 during da‐
ta log period has excee‐
ded the too low or too
high consumption limit
Cause
● Power failure during
Eeprom write
● Battery capacity low
● Meter has been
powered up
● Wrong sizing – flow
sensor too small
● Wrong settings for
output A
● Wrong settings for
output B
● Flowrate higher or
lower than expected
● Wrong parameter setup
Remedy
1. Reset checksum repair alarm via
parameter 560 and check data.
All data is checked after operation
(e.g., new flow calculation, writing
to the EEprom, etc.) with a
checksum control. If the checksum
result fails, data will not be
considered as valid and repair data
must be made.
2. If data is wrong or checksum error
reoccurs, replace the PCB board.
The alarm is active until the fault condition
is corrected.
Check calculated battery capacity (param‐
eter 510) versus battery alarm limit (param‐
eter 206) and replace batteries if necessary
Flow measurement and communication
stop, but display remains active as long as
power is available.
The alarm remains active until the fault con‐
dition is resolved.
Check meter sizing for actual installation
Alarm remains active until fault condition is
resolved.
Change volume per pulse to a higher value
- see Technical data (Page 96) for pulse
selection.
Reminder: basic version is limited to 50 Hz
maximum; advanced version to 100 Hz
maximum.
The alarm remains active until output pulse
rate drops below maximum pulse rate.
Change volume per pulse to a higher value
- see Technical data (Page 96) for pulse
selection.
Reminder: basic version is limited to 50 Hz
maximum; advanced version to 100 Hz
maximum.
The alarm remains active until output pulse
rate drops below maximum pulse rate.
Check data logger values and consump‐
tion limit.
Alarm remains active until it is manually re‐
set via parameter 209.
Operating Instructions, 11/2018, A5E03071515-AE
SITRANS MAG 8000