Functional Check Procedures - GE LOGIQ 400 Service Manual

Hide thumbs Also See for LOGIQ 400:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 1
4–2 FUNCTIONAL CHECK PROCEDURES
4–2–1 Basic Controls
Step
Check
1.
Connect a sector or convex
transducer to "Probe 2"
connector beneath keyboard.
2.
Turn ON Power toggle switch.
3.
Type USER ID and then press
return key as necessary.
4.
Press New Patient key.
5.
Press Top Menu Select key
twice.
6.
Select B section using
key.
7.
Press
8.
Select Dynamic Range section
using
9.
Change Dynamic Range up/
down using
10.
Press Top Menu Select key
twice.
11.
Rotate B/M Gain knob.
12.
Slide TGC potentiometers
(pots).
13.
Rotate Depth knob.
14.
Press Reverse key.
15.
Press Reverse key again.
Expected result
PATIENT ENTRY MENU should appear.
After a few seconds, the B mode screen should appear
as shown in ILLUSTRATION 4–1.
The top menu should appear on the CRT monitor screen
as follows:
The items displayed on the Top Menu are different de-
pending on the component options.
or
The B selection is displayed in reverse video.
key.
The B mode sub-menu should appear on CRT monitor
screen as follows:
The Dynamic Range selection is displayed in reverse
or
key.
video.
Image grows softer and harder depending on position.
or
key.
The menus on CRT monitor screen should disappear.
Image grows lighter and darker with rotation.
Image grows darker or brighter at depth equivalent to
pot's location.
The depth of image should be magnified/reduced.
The image reverses the left/right orientation.
The image reverses the left/right orientation.
LOGIQ 400 SERVICE MANUAL
Note
4–4
2127661
FUNCTIONAL CHECKS