M006-06
Rev. 01 – 05/2012
Aqua-Cleer
MFP 4–44 MD
Model 400-800-1200
1600-2200-2800-3300
USER'S
TECHNICAL MANUAL
MEDICAL DEVICE
0434
Culligan reserves the right to change the specifications given in this manual at any time.
CULLIGAN ITALIANA
S.p.A. -
Via Gandolfi, 6 - 40057 Cadriano di Granarolo E. BO (ITALY)
Tel. +39.051.6017.111 – fax +39.051.765.602
Company with UNI EN ISO 9001 certified quality system

Advertisement

loading

Summary of Contents for Culligan Aqua-Cleer MFP 4-44-MD 400

  • Page 1 USER’S TECHNICAL MANUAL MEDICAL DEVICE 0434 Culligan reserves the right to change the specifications given in this manual at any time. CULLIGAN ITALIANA S.p.A. - Via Gandolfi, 6 - 40057 Cadriano di Granarolo E. BO (ITALY) Tel. +39.051.6017.111 – fax +39.051.765.602...
  • Page 3 Any equipment operation fault not envisaged in this manual must be promptly reported to the Authorized Culligan  Service Center. Culligan declines any liability for the consequences of prolonged use of a unit having faults that were not reported. When necessary, the choice, dosing and handling of chemicals must be carried out according to the instructions given ...
  • Page 5: Table Of Contents

    CONTENTS GENERAL INFORMATION Page OPERATING LIMITS AND TECHNICAL CHARACTERISTICS ” DESCRIPTION OF SUPPLY “ LABELING AND IDENTIFICATION “ INSTALLATION “ DESCRIPTION OF ELECTRONIC PANEL KEYBOARD ” OPERATION FOR USER WATER PRODUCTION ” CONTROLS AND ADJUSTMENTS ” SANITIZATION “ ALARMS “ PROGRAMMING “...
  • Page 7: General Information

    Calcium Sulfate, Calcium Carbonate and Silica: specially developed and created by Culligan, on the basis of appropriate the limits specified refer to the need to prevent chemical analyses.
  • Page 8: Description Of Supply

    Table 1 Max. Max. Electrical connections Dimensions in mm Membrane Model flow flow Voltage Max. Approx. Min. flow max. feed MFP 4- rate, rate, shipment in V Max. absorpt rate, inlet water Frequency Heigh 44 MD treated drain weight input Width Depth water l/h...
  • Page 9 Photo No. 2A: Position of label on right side of frame Photo No. 4: “Reject Recirculation Flow” label Figure 2:Label on equipment Photo No. 5: “Reject Circulation Flow Adj.” label Meaning of data and symbols given on the identification label on outside of packing: manufacturer's mark, name and address of manufacturer,...
  • Page 10: Description Of Electronic Panel

    USER (SERVICE) WATER THE EQUIPMENT MUST ONLY BE INSTALLED PRODUCTION AUTHORIZED CULLIGAN SERVICE (unless otherwise specified, see fig. 3) CENTER TECHNICIANS. ATTENTION THE AQUA-CLEER MFP 4-44 MD SHOULD NOT When the main switch located on the side of the electronic panel is in...
  • Page 11: Controls And Adjustments

    1 again, regardless of the page displayed before the stop. N.B.: until the key is pressed To modify factory-set parameters, contact the authorized Culligan service center.. THERMAL ALARM In case of pump-motor unit overheating, the unit stops and the display N.B.: shows:...
  • Page 12 Leg. Decree 151-2005 ANY MAINTENANCE OPERATIONS CARRIED OUT BY THE USER WILL INVALIDATE ALL WARRANTIES AND RELIEVE CULLIGAN OF ATTENTION ANY LIABILITY. IN THIS WAY CULLIGAN ALSO GUARANTEES THE USE, WHEN NECESSARY, OF ORIGINAL REPLACEMENT PARTS, WHICH ONLY ONES ALLOWED CORRECT OPERATION OF THE EQUIPMENT.
  • Page 13 TABLES TABLE A (Table 201 Standard CEI EN 60601-1-2) GUIDE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS The AQUA-CLEER MFP 4-44 MD is designed to operate in the electromagnetic environment specified below. The AQUA-CLEER MFP 4-44 MD user must ensure that it is used in such an environment Emission test Conformity Electromagnetic environment...
  • Page 14 Activate the high-sensitivity differential switch. required for installation. Probable poor operation of the disconnecting switch installed on Contact the Authorized Culligan Service Center that carried the equipment electronic panel. out installation. Probable incorrect or only partial connection of the power cable Contact the Authorized Culligan Service Center that carried to the power supply terminals.
  • Page 15: Maintenance Schedule

    MAINTENANCE SCHEDULE date INLET signature CONDUCTIVITY CHECK date PRODUCT signature date PRODUCT signature date FLOW RATE REJECT signature CHECK date RECIRCULATION signature date INLET signature date CONDUCTIVITY PRODUCT CHECK signature date REJECT signature date S.D.I. TEST signature date ELECTRIC FUNCTIONS AND ALARMS EFFICIENCY CHECK signature date...
  • Page 16 Aqua-Cleer MFP4-44 - MD – User M006-06 – Rev. 01 – 05/2012...