Participant Registration Errors - Siemens Prisma 3T User Manual

Mri scanner
Table of Contents

Advertisement

Participant Registration Errors

1. You cannot confirm your registration due to the below error
 This happens when there are still active sequences in the Exam Card
 Go back to the Exam card and stop (red square) or Skip (>) the remaining
sequences in the queue
 This happens when a participant aborts a scan early or the group runs out of time
and stops scanning before their protocol is complete
 Also note, you will not be able to close the patient on the Exam card until the
sequences are stopped, skipped and/or deleted
2. The registration exam button is grayed out so you cannot click it
 This means you have forgotten to complete a required field
 Required fields are in bold
 NOTE: "Date of Birth" is bold, but we use "Age" instead – please do not input a
DOB – once you input the age a general DOB will be generated
 NOTE: Referring Physician is not bold but is required to access your data