Advertisement

5.2.5.

Setting menu display and operating procedures

1 : MONITORING MENU
2 : SETTING MENU
3 : MEASURING MODE
[ESC]
(Previous
Menu)
Test output will be
valid.
[→] : CONTINUE
[ESC]: CANCEL
Input password
PASSWORD
[ESC]
:0000
(Previous
[SET]SET,[ESC]CANCEL
Menu)
5:SET PARAMETERS
6:ZERO CALIBRATION
[ESC]
7:SPAN CALIBRATION
(Previous
8:ANGLE ROTATION
Menu)
[DN]
(Pre. menu)
9:LINEARIZ/CNDUCTVTY
10:ADDITIVES CORRECT
[ESC]
11:OTHERS
(Previous
Menu)
[DN]
(Pre. menu)
1 : READ PARAMETERS
2 : MEASURED VALUES
[ESC]
3 : SELF-DIAGNOSIS
(Previous
Menu)
Move the cursor to "2" with [→] key, and press [SET] key.
When [2: SETTING MENU] is selected, the density output
signal and the density display are hold in the simulated
values that have been set. This warning screen will
appear once before getting into [2: SETTING MENU].
After making sure that there is no problem, press the [→]
key to get into [2: SETTING MENU].
Press the [→] key to get into [2: SETTING MENU].
To get into [2: SETTING MENU], it is necessary to further
enter the password "8000". In the initial condition, the cursor
is on the forth digit. Therefore, press [DN] key 4 times to set to
"8" and press the [SET] key to get into [2: SETTING MENU]. An
incorrect password will cause the following message to
appear. Return to the screen at left by pressing any key and
then enter the correct password once again.
[UP]
(Next menu)
[UP]
(Next menu)
Move the cursor to the menu number with
[→] key, and press [SET] key.
 37 
6 F 8 A 0 5 2 1
PASSWORD ERROR
PUSH ANY KEY.