ABB Navigator 550 Commissioning Instructions page 7

Hydrazine wet-section
Hide thumbs Also See for Navigator 550:
Table of Contents

Advertisement

Blue
C
Red hydrazine
sensor connector
(upper socket)
Blue hydrazine
temperature sensor
connector
(lower socket)
Fig. 5 Setting up the sample flow
CI/AHM550–EN Rev. D
Red
E
D
b. Press the
key to select the Enter Configuration menu
and press the
key (below the
The Access Level page is displayed:
Access Level
Read Only
Calibrate
Advanced
Service
Back
c. Use the
key to scroll to the Advanced access level
and press the
key (below the Select prompt) to enter
the top level Configuration menus.
d. Use the
/
keys to scroll to the Calibrate page and
press the
key (below the Select prompt to enter the
Calibrate menus.
16. Use the
key to scroll to the Calibration Setup menu, press
the
key (below the Select prompt) to access the Standard
Solution menu.
17. Press the
key (below the Select prompt on the Standard
Solution page).
18. Enter the required concentration of the standard solution
using the
/
keys to scroll to the required numbers.
Press the
key (below the Next prompt) to save each
selected number. When the last number is highlighted, press
the
key (below the OK prompt) to set the value and exit
the page.
Note. If, during normal operation, the transmitter does
not display the expected hydrazine level, refer to the
Operating Instructions (OI/AHM550-EN) for
troubleshooting procedures.