  • Page 1 User Guide For Use With Ascensia ® ONTOUR Blood Glucose Meter and Ascensia ® ICROFILL Test Strips ® ®...
  • Page 2 Thank you for choosing the Ascensia Blood Glucose Monitoring System! We are proud to be your partner in helping you manage your diabetes. Our goal is to make this the simplest, most straightforward meter you will ever use. The Ascensia C System is intended for self-testing by people with diabetes and by healthcare professionals to monitor glucose concentrations in whole blood.
  • Page 3: Table Of Contents

    Table of Contents Your Testing Tools Your Meter Display Your Meter Features Symptoms of High or Low Blood Glucose Important Things to Know Before You Begin Getting Ready to Test Getting the Blood Drop Fingertip Testing Multiple Site Testing Testing Your Blood Error Codes and Symbols Control Solution Testing Setting the Date, Time, and Sound...
  • Page 4: Your Testing Tools

    Your Ascensia Up/Down Scroll Buttons Test Strip Port Your Ascensia Sample Tip: The end of the test strip where the blood is pulled in. Your Ascensia Re-setting End Release Button Lancet Holder Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Meter: ®...
  • Page 5: Your Meter Display

    Your Meter Display The picture below shows all the symbols that will appear on your meter display. When the meter is off, press and continue holding the ▲ or ▼ button. All segments will be displayed and remain on while holding the button down for 10 seconds. Repeat as needed to compare your meter to the display shown below.
  • Page 6: Your Meter Features

    Your Meter Features Programmed: Your Ascensia C Meter has been ONTOUR preset at the factory with the standard date, time, and meter features used where you live. To change these settings or use advanced meter features, such as meal markers, read through the Options section of your user guide.
  • Page 7: Symptoms Of High Or Low Blood Glucose

    Symptoms of High or Low Blood Glucose You can better understand your test results by being aware of the symptoms of high or low blood glucose; this will also help you decide what to do if your results seem unusual. Some of the most common symptoms are: High blood glucose (Hyperglycemia):...
  • Page 8: Important Things To Know

    Bayer Customer Service at the number below. Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Blood Glucose Meter is ONTOUR The bottle is especially ); results in mg/dL...
  • Page 9: Before You Begin

    Before You Begin Have all the materials you will need ready before you begin testing. This includes your Ascensia C the Ascensia M ICROFILL Lancing Device with Ascensia M ICROLET cets. You may also need Ascensia Solution to run a quality control check. Read your user guide and all instructional mate- rials provided in your meter kit before testing.
  • Page 10: Getting Ready To Test

    Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Getting Ready to Test: Read the Ascensia ICROLET Device insert for complete instructions. Remove the endcap from the Ascensia M ICROLET Device. Rotate (do not remove) the protective cap on a lancet ¼...
  • Page 11 Replace the device endcap. Remove a test strip from the bottle. Tightly close the bottle lid immediately after you have removed the test strip. NOTE: Check the expiry and dis- card dates. Make sure the test strip does not appear torn or damaged.
  • Page 12: Getting The Blood Drop

    Shallow Puncture Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Getting the Blood Drop: Fingertip Testing— The depth of the puncture depends upon the adjustable endcap setting you choose and the amount of pressure applied Deep to the puncture site. You should Puncture determine which combination works best.
  • Page 13: Multiple Site Testing

    Multiple Site Testing— Read the lancing device insert for complete instructions. Blood samples for glucose testing may be taken from sites other than your fi ngertips. Multiple Site Test- ing using blood from the forearm, palm, abdomen or thigh may give glucose results that signifi...
  • Page 14: Testing Your Blood

    Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Testing Your Blood: mmediately touch the tip of the test strip to the drop of blood. The blood is pulled into the test strip through the tip. Do not press the tip against Hold the tip of the test strip in the blood drop until the meter beeps.
  • Page 15 Expected Values: Blood glucose values will vary depending on food intake, medication dosages, health, stress or exercise. Consult your healthcare professional for the target value that is appropriate for you. Standard medical practice goals for a typical non-pregnant individual with diabetes are: •...
  • Page 16 Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) To remove the used lancet: Place the protective cap on a fl at surface with the Bayer logo facing down. Remove the endcap. (Do not use the same endcap on another person.) With the used lancet still in the lancing device, push the lancet needle completely into...
  • Page 17: Error Codes And Symbols

    Error Codes and Symbols: If you do not resolve the problem, contact Bayer Customer Service at 0845 600 6030 in the UK or 1 890 920 111 in Ireland. WHAT YOU SEE WHAT IT MEANS (will remain Low batteries on the display) (will fl...
  • Page 18 If you do not resolve the problem, contact Bayer Customer WHAT YOU SEE Glucose levels above 13.9 mmol/L may indicate a potentially serious medical condition. Glucose levels below 2.8 mmol/L may indicate a potentially serious medical condition. Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) WHAT IT MEANS WHAT YOU SHOULD DO Wrong test strip...
  • Page 19 Service at 0845 600 6030 in the UK or 1 890 920 111 in Ireland WHAT YOU SEE WHAT IT MEANS Control test • Test strip or result is out-of- control solution range (too high is past the expiry or too low). date or is past the discard date.
  • Page 20: Control Solution Testing

    Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Control Solution Testing: Use Ascensia M ICROFILL Solutions to practice on the sys- tem or to run a check to make sure the meter and the test strips are working properly. Using another control solution could present incorrect results.
  • Page 21 Gently rock the control bottle before opening to ensure it is mixed well. Squeeze a small drop of control solution on a clean nonabsorbent surface (such as a piece of wax paper). Do not apply control solution to the test strip directly from the bottle.
  • Page 22 If your control test result is out-of-range, see the “What You Should Do” recommendations on page 17. Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Compare your control test result with the Normal Control Range printed on the test strip bottle label or on the bottom of the test strip box.
  • Page 23: Setting The Date, Time, And Sound

    Setting the Date, Time, and Sound: You can set your meter to display different options. To get into set-up mode, press and hold M for three seconds. A complete display will briefly appear, then go to a display with the time. Setting the Time: The number in the hour position will be fl...
  • Page 24 Setting the Date: The date will now appear on the display. The year will be fl ashing. Press ▲ or ▼ to reach the current year. Press M to set. The month will now fl ash. Press ▲ or ▼ to reach the current month.
  • Page 25 Setting the 12 or 24 Hour Clock: 24H will now fl ash on the display. Press ▲ or ▼ to choose the 12H or 24H setting. This option lets you set the meter to display time as a 12 hour setting with AM and PM or as a 24 hour display (military time).
  • Page 26: Setting The Markers And Alarm

    Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Setting the Markers and Alarm: Markers and Alarm: You will now see three markers displayed on your meter along with the word “ Your Ascensia C ONTOUR give you the option of marking your test results.
  • Page 27: Using The Markers And Alarm

    Using the Markers and Alarm: If you have set your meter with the marker option on (see pages 21–24), your meter will display three markers next to your blood glucose test result. Pre-meal = Marks a test result that was taken before a meal.
  • Page 28 If you do not want to mark the test result: Press M or remove the used test strip to automatically save the test result in the meter memory. Press ▲ or ▼ until all three markers disappear from the display. Press M and remove the used test strip. Dispose of the used test strip and lancet properly to prevent injury or contamination to others.
  • Page 29 The Alarm: When the alarm sounds, the meter will automati- cally turn on, the alarm will sound 20 times, and the display will show the pre-meal test result you marked when you set the alarm. Pressing any but- ton will silence the alarm. You can now run another test by placing a test strip in the test strip port and following the...
  • Page 30 TO MARK RESULTS Pre-Meal Press ▲ or ▼ Alarm Press ▲ or ▼ Post-Meal Press ▲ or ▼ Logbook Press ▲ or ▼ Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) To Cancel an Alarm: Press M to enter the meter memory.
  • Page 31 Press▲ or ▼. The indicate the alarm has been turned off. Press M to set. YOU WILL SEE www.ascensia.co.uk will now WHAT IT MEANS Result is marked as a pre-meal test. You can now set an alarm. Alarm is set to go off in 2½ hours to remind you to run a post-meal test.
  • Page 32: Viewing Stored Results

    Using the Memory: With the meter off, press the M button on the meter to view stored test results. The 14 day average will appear on the display with the average for test results taken over the last 14 days. If you have set an alarm that has not sounded yet, you will also see a symbol.
  • Page 33 Press ▼ to review previous results. Previous results: Results will be shown starting with the most recent. Each result will show the date and time the test was taken as well as any symbols you chose to mark the test result. You can press ▲...
  • Page 34: Changing The Batteries

    Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) Changing the Batteries: Batteries should be replaced when the meter continually displays or the meter briefl y shows then turns off (See Error Codes and Symbols, page 15). Press fi rmly on the battery cover and slide in the direction of the arrow.
  • Page 35: Caring For The System

    Slide the battery cover back into place, lining up with the open slots, and close fi rmly. Keep batteries away from children. Lithium batteries are poisonous. If swallowed, imme- diately contact your physician or poison control center. Discard batteries according to your local environmental regulations.
  • Page 36: Transferring Results To A Computer

    Transferring Results to a Computer: IMPORTANT: The Ascensia C been tested or certified by Bayer for use with any software other than the Bayer diabetes management software. Bayer is not responsible for any erroneous results from the use of other software. Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland) You can transfer test results from...
  • Page 37: Technical Information

    Technical Information: Accuracy The accuracy of the Ascensia C ing System has been assessed in a correlation study conducted by healthcare professionals. In this study, 108 fresh capillary blood specimens were tested with the Ascensia C Glucose Monitoring System by 108 untrained people with dia- betes using three test strip lots.
  • Page 38: Specifi Cations

    Specifi cations: Test Sample: Whole blood Test Result: Referenced to plasma/serum glucose Sample Volume: 0.6 µL Measuring Range: Measuring Time: 5 seconds Memory Feature: Stores most recent 480 test results Battery Type: Two 3-volt lithium batteries (DL2032 or CR2032) Battery Life: Approximately 1000 tests (1 yr.
  • Page 39: Symbols Used

    Symbols Used: The following symbols are used throughout the product labeling for the Ascensia Monitoring System (meter packaging and labeling, test strip packaging and labeling and control solution packaging and labeling). Expiry date (use by last day of month) Batch code Temperature limitations Consult instructions for use In vitro diagnostic device...
  • Page 40 HealthCare instruments caused by the use of reagents other than reagents (i.e., Ascensia Strips) manufactured or recommended by Bayer HealthCare. Bayer HealthCare reserves the right to make changes in design of this instrument without obligation to incorporate such changes into previously manufactured instruments. Need Help?
  • Page 41: Warranty

    Bayer HealthCare has no knowledge of the perform- ance of the Ascensia when used with any test strips other than Ascensia Test Strips, and therefore makes no warran- ICROFILL ty of the performance of the Ascensia C when used with any test strips other than Ascensia...
  • Page 42: Service And Supplies

    Bayer Customer Service Representative. We have trained specialists to help you. Important: Speak to a Bayer HealthCare Representative before returning your meter for any reason. He/she will give you the information needed to get your problem handled correctly and effi...
  • Page 43 Supplies: When calling or writing for supplies be sure to include the number with the name of the replacement part or acces- sory item. Replacement Parts Part Number* 40030030 N1146C 40453276 Part numbers are subject to change without notice REFERENCES American Diabetes Association, www.diabetes.org Protection of Laboratory Workers from Occupationally Acquired Infections;...
  • Page 44 Chapel Lane, Swords, Co. Dublin, Ireland Bayer, the Bayer Cross, Ascensia, C are trademarks of Bayer. ACULANCE N1146C ONTOUR ICROFILL © 2006 Bayer HealthCare LLC Printed in UK Bayer Diagnostics Mfg. Ltd. Chapel Lane Swords, Co Dublin Ireland Phone—353 1 8132222 Fax—353 1 8132111...