Summary of Contents for Vector VI 200

  • Page 1 Weighing Indicator User manual...
  • Page 2 User Manual safety instruction For safety operation pls. follow the safety instruction. WARNING set. Calibrate, inspect and fix the the weighing indicator is prohibited by Non professional staff WARNING Pls. make sure the weighing display well earthing WARNING The indicator is electrostatic sensitive device, pls. power off during electrical connections, internal components touched by hand is prohibited, and please take the anti-static measure Page 1 of 27...
  • Page 3: Table Of Contents

    User Manual LIST 1. Summary ..................... 3  1.1 Main function ..................3  1.2 Technical parameter ................3  1.3 Outline and installation picture ............4  1.4 Battery ....................5  2. Installation and Calibration ................. 5  2.1 connection indicator with load cell ........... 5  2.2 communication interface ..............
  • Page 4: Summary

    User Manual 1. Summary 1.1 Main function 》General weighing: zero tare 》Animal weighing: Peak-hold. Data-hold, Auto-hold 》Accumulation 》Optional by RS232 》Low battery remind 》Power off automatically 1.2 Technical parameter 》Stimulating voltage:+3.3 VDC 》A/D converting speed: 10 SPS 》Load signal range: 0~12.8mV 》load capacity: it can connect 4 pcs 350 load cell at most Ω...
  • Page 5: Outline And Installation Picture

    User Manual 》Operation temperature: -10~40℃ -20~+60℃ 》 : Storage temperature 》optional power: 4V/4Ah rechargeable battery 110/220VAC 1.3 Outline and installation picture Page 4 of 27...
  • Page 6: Battery

    User Manual 1.4 Battery 1. when you use the internal battery first time,you should charge the battery 10-12 hours, to prevent low voltage resulted from self leakage of battery. 2. when the red battery light is on and flashes,it means low battery You should charge battery in time.
  • Page 7: Communication Interface

    User Manual Quick connector connection 2.2 communication interface PIN2-------TXD 5 4 3 PIN3------RXD PIN5 --------GND 8 7 6 3. Basic operation 3.1 Key and display Page 6 of 27...
  • Page 8 User Manual Weighing indicator display instruction LED display instruction Weighing data display Weight unit kg HOLD Hold the data Tare Display tare status Display net weight Page 7 of 27...
  • Page 9 User Manual Gross Display gross weight Display data keep still zero, indicating zero weight Battery Using battery Over Limit Within Limit Below Limit Total Accumulation Count Counting function Key’s function ON/OFF, EXIT AND SAVE SETTING LEFT RIGHT DOWN CONFIRM, GO TO NEXT STEP Page 8 of 27...
  • Page 10 User Manual Key symbol Key name Key function Work together with “on/off” enter and exit calibration ZERO Clear weight within zero range 1.At Gross mode, tare the loaded weight TARE 2.At Net mode, display gross weight after deduct tare HOLD Enter and exit “hold”...
  • Page 11: Power On

    User Manual 3.2 Power on Power on and indicator perform self-checking and go to weighing mode. 3.3 Zero setting Within zero range, press “zero”, indicator weighing is cleared. When Indicator is not stable, zero is unworkable. 3.4 TARE At the gross weight mode, if the weight is stable, pls. press ”Tare” key , the indicator will take the loaded weight as tare, and show net weight, At this time the gross mode will change to net mode.
  • Page 12: Total

    User Manual weighing indicator. Press “ HOLD” key again to exit the hold function C11=2 Data-hold Press” HOLD” key, the Hold light is on, and show the data on the weighing indicator. Press “ HOLD” key again to exit the hold function C11=3 Auto-hold If the weight on the scales above 20d and keep stable, the indicator will show the data for 6 seconds and the “...
  • Page 13: Times High Resolutions

    User Manual Check the total weight operation: Press “SET” hold it then press “TOTAL” At the same time, display ”n**”, (accumulating times) then display total weight. There are 8 data totally. It shows the first 4 digital. then the last 4 digital For example, the first 4 digital is”0012”, the last 4 digital is”34,56”...
  • Page 14: Print Function

    User Manual weight is below than the low limit, the “LO” light will on.. when the weight is within the limit, the “OK” light is ok. 3.9 Print function When the data is stable, connection with printer, it will be printed after press“...
  • Page 15: Step Of Calibration Operation

    User Manual down the “span” Press still and then press at the same time , you will enter C01-C39 setting. 4.2. Step of calibration operation: According to the second method which can enter setting menu, C01-C39 C01 UNIT [C01 ] Press 1] unit is kg Press , go to next step C02 Set decimal digits...
  • Page 16 User Manual [C03 ] Press 1] d=1 2] d=2 5] d=5 10] d=10 20] d=20 50] d=50 Press , go to next step C04 Maximum capacity For example:max weighing 100kg: set [0100.00] Press , go to next step C05 Zero calibration option:0=non-calibration zero 1=need calibration zero calibration zero please choose 1 and ensure scale is empty and “stable”...
  • Page 17: Application Parameter Setting

    User Manual As bellows: [0080.00] [CAL …… [0080.00] [CALEnd] count down over, indicator shows loaded weight , loading calibration finish. If you want to set application parameter. Press If you want to exit. Press C07 Default parameters setting [C07 0] non-restore default parameters [C07 1] restore default parameters Note: after the above parameters setting finish, please do not set default parameters often, avoid the original setting parameters lost.
  • Page 18: Communication Setting

    User Manual C10 Power saving setting [C10 0] close backlight [C10 1] close backlight after 1 minute [C10 2] always backlight C11 Hold [C11 0] No Hold function [C11 1] Peck hold [C11 2] Data hold [C11 3] Auto-hold [C11 4] Animal weighing C12 Hold time ( if you choose C11=4, you can set the time )
  • Page 19: Application Setting

    User Manual [C19 2400bit/s [C19 4800bit/s [C19 9600bit/s 4.5 Application setting C20 Manually Zero [C20 00] no Manually Zero [C20 01] Manually Zero range ±1% Max. capacity [C20 02] Manually Zero range ±2% Max. capacity [C20 04] Manually Zero range ±4% Max. capacity [C20 10] Manually Zero range ±10% Max.
  • Page 20 User Manual [C22 4.0] ±4.0d [C22 5.0] ±5.0d C23 Zero tracking time [C23 0] No zero tracking [C23 1] 1 second [C23 2] 2 seconds [C23 3] 3 seconds C24 Overload range [C24 09] Over 9d than max. capacity C25 Negative display [C25 00] Less than -9d [C25 10]...
  • Page 21: Exit Setting

    User Manual [C28 5] High dynamic filter C29 Noisy filter [C29 0] Close noisy filter [C29 1] Low noisy filter [C29 2] Medium filter [C29 3] High filter 4.6 Exit setting For example [C10 1], Press , confirm it then press to exit and save it 5.Output data format 5.1Computer continuous sending format...
  • Page 22: Big Display Continuous Sending Format

    User Manual weight mode, GS=gross mode, NT=net mode weight of positive and negative, “+” or ” –“ measurement unit, “kg” or “lb” Data: weight value, including decimal point carriage return line feed 5.2 Big display continuous sending format Output continuous format State A Bits0,1,2 Decimal point position XXXXXX0 XXXXXXX XXXXX.X...
  • Page 23: Serial Interface Reception Command

    User Manual State B BitsS function Bits0 gross=0, net=1 Bits1 symbol:positive =0,negative =1 Bits2 overload(or lower zero)=1 Bits3 dynamic=1 Bits4 unit:lb=0, kg=1 Bits5 Constant 1 Bits6 Constant 0 State C Bit2 Bit1 Bit0 unit Kg or lb Bit 3 printing=1 Extend Bit 4 display=1...
  • Page 24: Print Output Format

    User Manual Command name role Tare off command Save and clear tare Zero command Zero the gross weight Print command Print the weight Read gross/ Read gross/net weight weight 5.4Print output format 004 (NO.) G.W: 8.88kg (gross,example for two decimal point) T.W: 2.88kg (tare)...
  • Page 25: Maintenance

    User Manual 6. Maintenance 6.1 Regular Error and maintain method Error Reason instruction Solution 1. the loaded weight 1.decrease loaded weight excess overload range of check load cell max. capacity connection Display wrong connection 3. checking load cell:check UUUUUU with load cell or no input and output resistance connection with it.
  • Page 26: Daily Maintenance

    User Manual during calibration, the Added weight least added weights 10%of Max. capacity, ERR2 enough Recommend the weights is 60-80% the Max. capacity during calibration, input 1..Check connection single is negative. correct or not. Check load cell is damaged ERR3 or not.
  • Page 27: Restore Default Parameters

    User Manual 6.3 Restore default parameters Enter setting menu, set C07= 1,press then press exit saving setting, all parameters will be back to default setting. Note:Pls. do not restore default parameter easily if you are not professional and have not scale calibration. Default parameter form parameter instruction Default value...
  • Page 28 User Manual Date Time Serial interface data output methord Serial interface Baud rate 3=9600 Manual zero setting Initical zero setting Automatic zero tracking range Automatic zero tracking time Verload range Negative display range Standstill time Standstill range Dynamic filter Noisy filter C30~C40 Reseverd menu Page 27 of 27...