Note: When the calibration procedure is
completed, the CAL tag is turned on.
• To quit the procedure and return to the last
calibration data, after entering the calibration
mode press the ON/OFF button.
The secondary LCD displays "ESC" for 1
second and then the meter returns to the
normal measurement mode.
• To reset to the default values and clear a
previous calibration, press the SET/HOLD
button after entering the calibration mode and
before the first point is accepted.
The secondary LCD displays "CLR" for 1
second, the meter resets to the default
calibration and the CAL tag on the LCD turns
off.
SETUP
Setup
mode
allows
the
selection
temperature (°C or °F) and pH buffer set for
calibration. To enter the Setup mode, press
the ON/OFF button until "CAL" on the
secondary LCD is replaced by "TEMP" and the
current temperature unit (e.g. TEMP °C).
Then:
• for °C/°F selection: use the SET/HOLD
button; then press the ON/OFF button once to
enter the buffer set selection or twice to return
to the normal measurement mode.
• to change the calibration buffer set: after
setting the temperature unit, press ON/OFF
once and select the buffer set ("pH 7.01
BUFF" or "pH 6.86 BUFF" for NIST) by using
the SET/HOLD button. Press ON/OFF to
return to the normal measurement mode.
ELECTRODE REPLACEMENT
• Remove the protective cap and unscrew the
plastic ring on the top of the electrode.
• Pull out the MI58P electrode and replace it
with a new one.
• Make sure the gaskets are in place before
screwing back the ring.
BATTERY REPLACEMENT
When the batteries become weak, the battery
symbol on the LCD will light up to advise that
only a few hours of working time is remaining.
The meter is also provided with BEPS (Battery
Error Prevention System), which avoids any
erroneous readings due to low battery level by
automatically switching the meter off.
It is recommended to replace the batteries
immediately.
of
To replace the batteries unscrew the battery
compartment cap and replace all four 1.5V
batteries while paying
attention to their
polarity.
Make sure the gasket is in place before
screwing back the cap.
Batteries should only be replaced in a
nonhazardous area using the battery type
specified in this instruction manual.
ACCESSORIES
MI 58P
Replaceable electrode for pH 58
M10004B
pH 4.01 buffer, 20 ml sachet, 25 pcs.
M10007B
pH 7.01 buffer, 20 ml sachet, 25 pcs.
M10010B
pH 10.01 buffer, 20 ml sachet, 25
pcs.
MA9004
pH 4.01 buffer, 230 ml bottle
MA9006
pH 6.86 buffer, 230 ml bottle
MA9007
pH 7.01 buffer, 230 ml bottle
MA9009
pH 9.18 buffer, 230 ml bottle
MA9010
pH 10.01 buffer, 230 ml bottle
MA9015
Electrode storage solution, 230 ml
MA9016
Electrode cleaning solution, 230 ml
MA9020
ORP test solution (200/275 mV), 230
ml bottle
M10000B
Electrode rinse solution, 20 ml sachet,
25 pcs.
WARRANTY
This instrument is warranted against defects in materials
and manufacturing for a period of two years from the date
of purchase. Probe is warranted for 6 months. If during
this period the repair or replacement of parts is required,
where the damage is not due to negligence or erroneous
operation by the user, please return the parts to either
distributor or our office and the repair will be done free of
charge. Damages due to accidents, misuse, tampering or
lack
of prescribed maintenance are not covered.
Milwaukee/Martini instruments reserves the right to make
improvements in design, construction and appearance of
its products without advance notice.
SPECIFICATIONS
RANGE
-2.00 to 16.00 pH
±1000 mV (ORP)
-5.0 to 60.0°C / 23.0 to 140.0°F
RESOLUTION
0.01 pH / 1 mV
0.1°C / 0.1°F
ACCURACY (@25°C)
±0.05 pH / ±2 mV
±0.5°C / ±1°F
TYPICAL EMC DEVIATION
±0.02 pH
±2 mV
±0.3°C / ±0.6°F
TEMPERATURE COMPENSATION
Automatic for pH, from -5 to 60°C
CALIBRATION, pH Automatic, 1 or 2 points
with 2 sets of memorized buffers(pH 4.01,
7.01, 10.01 or 4.01, 6.86, 9.18)
CALIBRATION, ORP Factory calibrated
PROBE
Replaceable MI 58P
ENVIRONMENT -5 to 50°C; 100% RH max.
BATTERY TYPE 4 x 1.5V; IEC LR44, A76
BATTERY LIFE approx. 250 hours of use
AUTO-OFF
after 8 minutes of non-use
DIMENSIONS
200 x dia 38 mm
WEIGHT
100 g