Suggested Location Of Sampling Point - Bacharach 3015-4286 Installation & Operating Instructions Manual

Ammonia gas monitor - single zone
Table of Contents

Advertisement

Suggested Location of Sampling Point

At the point of an ammonia gas leak the gas is nearly pure. As ammonia gas is dispersed into the air, the gas
molecules diffuse causing a dilution of the original concentration. The AGM-SZ measures the gas concentration
at the point the sample is collected. Therefore, if the termination of the collection line is not at the exact point of
the gas leak, then the monitor will read a diluted mixture of ammonia gas and air.
It should be noted that when ammonia gas is cold, it is heaver than air and settles below the leak point. When
the gas warms to room temperature, it becomes lighter than air and tends to collect above the leak point.
Consequently, the sampling point should be located as close as possible to the source of a potential leak. If this
is impractical, then the alarm set points should be adjusted to compensate for the dilution of the ammonia gas.
The AGM-SZ should be mounted outside of the mechanical room or just inside a doorway to the room. This is
the "split architecture design" for safety of the operator. The monitor should be readily accessible for easy visual
monitoring and servicing. The total length of the sample line + exhaust line can be up to 500 ft (152.4 m). If the
area around the monitor is not well ventilated, then an optional exhaust line and purge line can be run to an
outside location. The total length of the exhaust line + the purge line cannot exceed 200 ft (61.0 m).
It may be necessary to perform a "smoke" test of the mechanical room to determine the best monitoring point.
The smoke test would provide the pattern of air currents present in the room.
AGM-SZ Refrigerant Gas Leak Monitor Mechanical Room Placement
4
Instruction 3015-4286

Advertisement

Table of Contents
loading

This manual is also suitable for:

Agm-sz

Table of Contents