Hart® Connections; Power Supply From The Calibrator - GE Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
6.3
HART® Connections
Before setting-up the electrical connections between the
HART® device and the Druck DPI 620 Genii get the correct
connection scheme (Ref: DASHBOARD >> HELP).
6.4

Power Supply from the Calibrator

A 24 V or 28 V loop power can be supplied using the CH2
Current (24V) measure function.
In the example that follows, the Druck DPI 620 Genii supplies
the loop power and a 250 Ω HART® resistor.
06 June 2013
Figure 6-1
K0541 issue 1
Druck DPI620 Genii
Page 94 of 175