Installation; Assembly And Wiring; Figure 2-1 Cordset Connection And Wiring - Honeywell DirectLine DL421 User Manual

Sensor module for durafet ii, durafet iii, meredian ii, and hpw7000 ph electrodes
Table of Contents

Advertisement

2.1 Assembly and Wiring

Depending on the customer selected output cable options, the DirectLine can be
wired to an appropriate 16-42 Vdc source using 2 different methods:
1) cordset
2) field wiring with customer supplied cable.
Refer to Section 7 for wiring for CE Mark applications.
2.1.1 Cordset
Chart Recorder,
Controller, etc.
(Use 250 ohms
for 1-5 V signal)
2/04

2. Installation

Power Supply
Minimum: 16 V @ 250 ohms (max.)
Maximum: 42 V @ 1400 ohms (max.)
_ _
_
Load
Resistance
+
Black (–)
*Blue wire not used

Figure 2-1 Cordset Connection and Wiring

DirectLine® DL421 Sensor Module User Manual
+ +
Brown (+)
Shield
Wire
(Ground)
*
Installation
Cordset 6m (19.7')
Electronics
Module
5