66103A DC Power Supply
Serial Numbers: Serial Numbers: MY40005783/MY40006268
The instrument will not communicate or will not complete self test.
Parts Required:
P/N
1200-1274
SOCKET-IC- 68-CONT PLCC
1821-1479
IC 16-Bit MCU 16 MHZ CMOS 68PLCC
6610X-61022
PCA-TSTD PWRMESH
IMPORTANT
Consult the Service Manual for the 6610XA Power Modules p/n 5959-3364. The information in the Service Manual should be
used during the disassembly and reassembly see page 57. The Service Manual can be viewed on line at the web address described
below.
When doing this Modification the following steps are required.
1.
Follow proper procedures to eliminate ESD damage to the instrument and components.
2.
The AC line MUST be turned OFF at the source and disconnected from the instrument.
3.
The correct hand tools should be used for disassembly and assembly procedures.
http://cp.literature.agilent.com/litweb/pdf/5959-3364.pdf
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION
[[]] ON SPECIFIED FAILURE
CATEGORY:
X AGREEABLE TIME
LOCATION
x
CUSTOMER INSTALLABLE
CATEGORY:
x
ON-SITE
x
SERVICE CENTER
[[]] CHANNEL PARTNER
PRODUCT'S SUPPORT LIFE
AVAILABILITY:
AUTHOR: CP
ADDITIONAL INFORMATION: This Service Note was originally posted as a Modification Available Service Note.
© AGILENT TECHNOLOGIES, INC. 2008
PRINTED IN U.S.A.
November 3, 2008
Rev. 15
R
E
C
O
M
M
E
N
D
E
D
R
E
C
O
M
M
E
N
D
E
D
R
E
C
O
M
M
E
N
D
E
Description
D
Qty.
1
1
STANDARDS
LABOR: 1.5 Hours
SERVICE
[[]] RETURN
INVENTORY: [[]] SCRAP
X SEE TEXT
NO CHARGE AVAILABLE UNTIL: (Always)
PRODUCT LINE: SP
66103A-08A
N O T E
Supersedes:
66103A-08
USED
[[]] RETURN
PARTS:
x
SCRAP
[[]] SEE TEXT
Page 1 of 5

Advertisement

loading

Summary of Contents for Agilent Technologies 66103A

  • Page 1 66103A-08A S E R V I C E N O T E Supersedes: 66103A-08 66103A DC Power Supply Serial Numbers: Serial Numbers: MY40005783/MY40006268 The instrument will not communicate or will not complete self test. Parts Required: Description Qty. 1200-1274 SOCKET-IC- 68-CONT PLCC...
  • Page 2: Special Note

    Page 2 of 5 66103A-08A SPECIAL NOTE: When a 6610XA comes in for calibration or repair and the instrument has a serial number with in the range shown above the socket may need to be replaced? To identify if the socket is defective see Figure 2 below for specific details.
  • Page 3 Page 3 of 5 66103A-08A Figure 2 This figure should be used to identify GOOD sockets and a potential defective socket. The 16-Bit MCU 16 MHZ 68PLCC CMOS IC (P/N 1821-1479) is NOT pictured. Manufacture’s Logo Manufacture’s Logo Manufacture’s Logo “A”...
  • Page 4 Page 4 of 5 66103A-08A “A” & “B” Discipline Discipline A – Permanent Corrective Action Plan “The socket shown in the pictures are for reference ONLY” The condition of the recessed beams has been corrected by a design enhancement implemented in production. This design enhancement increased the spring height dimension by 0.04mm and decreased the spring height tolerance by...
  • Page 5 Special Note The following Service Notes described below may have the same problem as described above. 66101A-09A, 66102A-08A, 66103A-08A, 66104A-09A, 66105A-09A, 66106A-08A, 66000A-06A Related Service Notes The following Service Notes may have the same problem “The instrument will not communicate or will not complete self test.”...