5.3 Measure
pressure/vacuum
5.3.1 Procedure overview
3
2
1
TAP
A
3
B
C
D
1
E
1
TAP
[EN] English - K0457
When the pressure calibrator assembly is complete
(Section 5.2), use the calibrator to measure the pressure or, if
applicable, the vacuum in the pressure station; for example: to
calibrate a gauge or indicator.
To use the pressure calibrator, complete these procedures:
•
Set the calibrator functions you want to use on the display:
a. Pressure function (P1): item
see Section 5.3.2
b. CH1: Channel 1 electrical function (input or output);
refer to the user manual - K0449
c. CH2: Channel 2 electrical function (input or output);
refer to the user manual - K0449
d. other functions (maximum: 6 functions);
refer to the user manual - K0449
•
If necessary, change the Units for the function: item
see Section 5.3.3
•
If necessary, set a Utility for the function: item
a. Max/Min/Avg; refer to the user manual - K0449
b. Switch Test; refer to the user manual - K0449
c. Leak Test; see Section 5.3.4
•
If necessary, change the Settings for the pressure function:
item
D
a. Process (Tare, Alarm, Filter, Flow, Scaling);
refer to the user manual - K0449
b. Leak Test (Only when the Utility is set); see Section 5.3.4
c. Zero; see Section 5.3.5
•
If necessary, change the Process for the CH1 and/or CH2
input functions: item
This includes: Tare, Alarm, Filter, Flow, Scaling; refer to the
user manual - K0449
A
;
E
Pressure calibrator operation (DPI 620) 5-3
Issue 1
B
;
C