Read/Reset Max Min Values; Write Cable Resistance - ABB 691HT Operating Instructions Manual

Hand held communicator intrinsic safety version
Table of Contents

Advertisement

7
In Write Protect Mode
the transmitter is in write protect mode
9
CJC too low
Input value for CJC is lower than expected
10 RTD multiplication factor too high
Multiplication factor in input is too high
11 CJC too high
Input value for CJC is higher than expected
12 RTD multiplication factor too low
Multiplication factor in input is too low
COMMAND 143: WRITE CABLE RESISTANCE
Command-Specific Response Codes
2
Invalid Selection
the selection does not correspond with the
transmitter type
5
Too few bytes received
the number of bytes received is less than
expected
6
Transmitter specific command error
the transmitter generates an unexpected error
7
In Write Protect Mode
the transmitter is in write protect mode
COMMAND 147: WRITE CURRENT OUTPUT RANGE
VALUES
Command-Specific Response Codes
2
Invalid Selection
the selection does not correspond with the
transmitter type
5
Too few bytes received
the number of bytes received is less than
expected
6
Transmitter specific command error
the transmitter generates an unexpected error
7
In Write Protect Mode
the transmitter is in write protect mode
9
100% Current value too high
the value defined for current is out of the high
limits
10 0% Current value too low
the value defined for current is out of the low
limits
11 Upper Current Limit value too high
the value defined for upper current limit is out of
the high limits
12 Lower Current Limit value too low
the value defined for lower current limit is out of
the low limits
COMMAND 151: WRITE SENSOR ERROR VALUES
Command-Specific Response Codes
2
Invalid Selection
the selection does not correspond with the
transmitter type
5
Too few bytes received
the number of bytes received is less than
expected
6
Transmitter specific command error
the transmitter generates an unexpected error
- 48 -
7
In Write Protect Mode
the transmitter is in write protect mode
9
Shorted sensor value too high
the value defined for shorted sensor is too high
10 Shorted sensor value too low
the value defined for shorted sensor is too low
11 Broken sensor value too high
the value defined for broken sensor is too high
12 Broken sensor value too low
the value defined for broken sensor is too low
COMMAND 153: READ/RESET MAX MIN VALUES
Command-Specific Response Codes
2
Invalid Selection
the selection does not correspond with the
transmitter type
5
Too few bytes received
the number of bytes received is less than
expected
COMMAND 154: RESET DAC TRIM
Command-Specific Response Codes
7
In Write Protect Mode
the transmitter is in write protect mode
TRANSMITTER SPECIFIC COMMANDS
( for KST TEMPERATURE Transmitters)
COMMAND 131: WRITE STATIC DATA - CONFIGU-
RATION
Command-Specific Response Codes
7
In Write Protect Mode
the write protect link on the transmitter is "on"
8
Warning: 4/20 points set to selected sensor limits
11 Invalid Sensor Type Code
Wrong Type of Sensor specified
12 Invalid Number of Wires Code
Invalid number of wires specified for the
specific sensor
COMMAND 133: WRITE LOWER TRIM POINT
COMMAND 134: WRITE UPPER TRIM POINT
Command-Specific Response Codes
3
Passed Parameter too large
4
Passed Parameter too small
The maximum change is 5% of the span
7
In Write Protect Mode
The Write protect link on the transmitter is "on"
10 Calibration Location not set to User
The Factory set trim parameter cannot be chan-
ged (see Sensor Trimming option )
11 Excess of correction attempted
12 Invalid Calibration Point Units Code
The calibration units are not proper for the type
of sensor
13 Trim span too small
The trim span is smaller than the span limits