66
Specifications

Physical

Dimensions
Weight
Operating Temperature
Storage Temperature
(without battery)
Humidity
Altitude
Shock
Vibration
Water resistance
Drop Test

Electrical Isolation

Power
External Electrical Connections
Risk Current Category

Data Management

Event Documentation
Internal Memory Capacity
Quick Report
Communication
8 1/4" x 6 7/8" x 2 3/4" (210 x 175 x 70 mm)
3.5 (1.6 kg) pounds with battery
32ºF to 122ºF (0ºC to 50ºC)
-22ºF to 158ºF (-30ºC to 70ºC)
Up to 95% (non-condensing)
Up to 15,000 feet (-150 to 4570 m)
Mil Std 810E method 516.4, procedure 1 (30G, 6-9ms pulse, 1/2 sine each axis)
Mil Std 810E method 514.4, category 1
IEC 60529 IPX4
Mil Std 810F method 516.5 Procedure IV (1 meter drop onto any edge, corner, or surface)
Unit operates on internal battery only
No external devices are attached to the unit
Internally powered equipment with defibrillator-proof BF type patient applied part (as
per definition of IEC 60601-1 standard)
Internal
1MB: 250 4-sec ECG samples or 3000 time stamped events
Treatment Summary, Event Log, Test Log
RS-232 via Infrared (IR) port to PC or printer