6-5-5. Other details

(1) Print at Zero
When moisture is not detected and result is
"0μg", select "Use" or "No use".
The default is "Use".
(2) Potential
When using reagents of different titration
speed, set electrolysis control start
potential. The standard value is 100mV.
When using electrolysis solution of fast
reaction speed, set larger values than the
standard value. When using electrolysis
solution of slow reaction speed, set smaller
values than the standard value. When using
RdH
anode solution, set it to 200mV.

6-5-6. Memory initialize

POINT
By setting the unit to default by memory initialization, all mode switch settings are set to
default. (Refer to 6-5-1. Mode change.)
In addition, all measurement data are canceled. To return to the condition before memory
initialization, run export previously and back up settings. (Refer to 7-2. USB memory.)
Initialize all memories and reset them to default.
Select "Yes" and press [Enter] key.
Other details
01 Print at Zero : Use
02 Potential
Select
key, Set Enter key.
Memory initialize
01 No
02 Yes
6-19