ABB SM500F User Manual page 144

Field mountable paperless recorder
Hide thumbs Also See for SM500F:
Table of Contents

Advertisement

SM500F
Field mountable paperless recorder
Relays
Number of relays
1 as standard, 2 optional, (1 Module)*
* When using dual relay option boards, high voltages above 120 V AC must not be connected alongside low voltage DC
supplies.
Type and rating
Relay type
Voltage
Current
Loading (non-inductive)
For dual relay boards, only the following permutations are permitted.
30 V DC / 30 V DC
120 V AC / 30 V DC
120 V AC / 120 V AC
240 V AC / 240 V AC
240 V AC / 120 V AC
AC supplies must be on the same phase.
2-Wire Transmitter Power Supply (optional)
Number
2 isolated supplies
Voltage
24 V DC nominal
Drive
22 mA (each supply)
Ethernet Module (optional)
Physical medium
10BaseT
Protocols
TCP/IP, FTP (server), HTTP, SMTP, Modbus TCP (Client + Server)
FTP server functions
Directory selection & listing
File upload/download
4 independently configurable users with full or read-only access
Web server functions
Operator screen monitoring/selection. Remote monitoring of recording channels, analog/digital signals, alarms,
totalizers and archiving
SMTP client compatibility
Compatible with MS Exchange versions up to and including MS Exchange 2003
142
Selectable NO/NC
250 V AC
30 V DC
5 A AC
5 A DC
1250 VA
150 W
8 Specification
IM/SM500F Rev. Z